lopital Nemo Cleaning And Disinfection Protocol page 35

Shower -toilet chairs, shower trolleys, dressing tables, rising aids
Table of Contents

Advertisement

Cleaning protocol
1. Before cleaning and disinfecting, unplug the product if applicable.
Cleaning
2. Clean the trolley/dressing table with a damp cloth with warm water and an institution-approved
neutral detergent. A soft brush can be used to remove stains and stubborn dirt.
3. Wipe the entire trolley/dressing table from top to bottom. Do not use a dripping wet cloth. For
easy access to all parts of the trolley/dressing table, adjust the trolley/dressing table to both the
highest and lowest position.
4. Pay extra attention to the following surfaces:
Trolley surface
-
Side rails
-
Manual operation (with electrically powered trolley)
-
Foot pedal (with hydraulic powered trolley)
-
Brake pedals
-
On/off switch (with electrically powered trolley)
-
Emergency lowering button and emergency stop (with electrically powered trolley)
-
Back adjustment button (with hydraulically powered trolley)
-
Wheels
-
You can also use a cleaning unit/cleaning system to clean your trolley/dressing table. The temperature in
the system should be a maximum of 55°C.
Disinfecting
5. Repeat the cleaning protocol - Cleaning after reading "Applying common cleaning/disinfectants to
Lopital products" on page 3 of this document. Follow the manufacturer's directions for proper use
of the disinfectant.
6. After disinfecting the trolley/dressing table, wipe the trolley/dressing table with a damp cloth and
clean water to remove any residue from the disinfectant.
34 | Cleaning and disinfection protocol