T +31 (0) 13 5239300 www.lopital.com info@lopital.nl All of the aids produced by Lopital fulfil the following norms: MDR (EU) 2017/745, NEN-EN-ISO 14971, NEN- EN 12182, ISO17966, IEC60529, IEC60601-1, IEC60601-1-2, IEC6060-1-6, IEC62366, CAN/CSA-C22.2 60601-1, ANSI/AAMI ES60601-1, UNI-CEI-EN-ISO 13485, NEN-EN-ISO 9001, NEN-EN-ISO 14001...
Lopital recommends cleaning and disinfecting the products when they are disposed of as waste. All parts that are replaced or removed can be returned to Lopital. It is also possible to offer these parts to the technician for destruction. We will ensure that all parts are processed in an environmentally friendly manner.
Applying Common Cleaning/disinfectants to Lopital Products Chemical class Active component Cleaning Manufacturer Must not be used on the agents/Disinfectants* following parts: Quaternary Didecyl dimethylammoniumchloride = 8.704% 9.0-10.0 Virex II (256) Johnson/Diversey Foot board of the Lotus rising aid ammonium Alkyldimethylbenzylammoniumchloride = 8.19%...
All necessary precautions should be taken to remove all stains and dirt. The cleaning and disinfection method described on the following pages applies specifically to Lopital products and accessories. This method is designed to save time and prevent infections more efficiently.
Nemo - Shower-toilet Chair 51005100 The Nemo must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
Page 7
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Application of common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
2017/745 a chair must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
Page 9
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Application of common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
2017/745 a chair must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
Page 11
Disinfecting 5. Repeat the cleaning protocol - Cleaning after reading “Application of common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 6. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
2017/745 a chair must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
Page 13
Disinfecting 5. Repeat the cleaning protocol - Cleaning after reading “Application of common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 6. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
2017/745 a chair must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
Page 15
Disinfecting 5. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 6. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
Tango – Shower-toilet chair 51005700 The Tango must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
Page 17
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
Mylo - Shower-toilet chair 51005701 The Mylo must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
Page 19
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
Tango XL – Shower-toilet chair 51005705 The Tango XL must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
Page 21
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
Tango XXL – Shower-toilet chair 51005800 The Tango XXL must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
Page 23
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
Timo – Shower(-toilet) chair 54006020, 54006021, 54006024, 54006030, 54006034, 54006035 The Timo must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products.
Page 25
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
Cleo – Shower(-toilet) chair 54006040, 54006050 The Cleo must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
Page 27
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
2017/745 the Marina basic / Marina deluxe must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the trolley thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
Page 29
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the trolley, wipe the trolley with a damp cloth and clean water to remove any residue from the disinfectant.
2017/745 the Marina must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the trolley thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
Page 31
Disinfecting 5. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 6. After disinfecting the trolley, wipe the trolley with a damp cloth and clean water to remove any residue from the disinfectant.
2017/745 the trolley/dressing table must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the trolley/dressing table thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
Page 33
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the trolley/dressing table, wipe the trolley with a damp cloth and clean water to remove any residue from the disinfectant.
2017/745 the trolley/dressing table must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
Page 35
Disinfecting 5. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 6. After disinfecting the trolley/dressing table, wipe the trolley/dressing table with a damp cloth and clean water to remove any residue from the disinfectant.
2017/745 the Luxal must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the trolley thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
Page 37
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the trolley, wipe the trolley with a damp cloth and clean water to remove any residue from the disinfectant.
Lotus – Rising aid 59008100 The Lotus must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
Page 39
Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the Lotus, wipe the Lotus with a damp cloth and clean water to remove any residue from the disinfectant.
Page 40
For more information, visit our website: www.lopital.com Lopital Nederland B.V. Laarakkerweg 9, 5061 JR Oisterwijk, Postbus 56, 5060 AB Oisterwijk Tel +31 (0)13 5239300, E-mail info@lopital.nl, www.lopital.com 39 | Cleaning and disinfection protocol REI-DES-01 05-2021...