lopital Nemo Cleaning And Disinfection Protocol

lopital Nemo Cleaning And Disinfection Protocol

Shower -toilet chairs, shower trolleys, dressing tables, rising aids

Advertisement

Cleaning and disinfection
protocol
Shower(-toilet) chairs
Shower trolleys
Dressing tables
Rising aids

Advertisement

Table of Contents
loading

Summary of Contents for lopital Nemo

  • Page 1 Cleaning and disinfection protocol Shower(-toilet) chairs Shower trolleys Dressing tables Rising aids...
  • Page 2: Table Of Contents

    T +31 (0) 13 5239300 www.lopital.com info@lopital.nl All of the aids produced by Lopital fulfil the following norms: MDR (EU) 2017/745, NEN-EN-ISO 14971, NEN- EN 12182, ISO17966, IEC60529, IEC60601-1, IEC60601-1-2, IEC6060-1-6, IEC62366, CAN/CSA-C22.2 60601-1, ANSI/AAMI ES60601-1, UNI-CEI-EN-ISO 13485, NEN-EN-ISO 9001, NEN-EN-ISO 14001...
  • Page 3: Introduction

    Lopital recommends cleaning and disinfecting the products when they are disposed of as waste. All parts that are replaced or removed can be returned to Lopital. It is also possible to offer these parts to the technician for destruction. We will ensure that all parts are processed in an environmentally friendly manner.
  • Page 4: Applying Common Cleaning/Disinfectants To Lopital Products

    Applying Common Cleaning/disinfectants to Lopital Products Chemical class Active component Cleaning Manufacturer Must not be used on the agents/Disinfectants* following parts: Quaternary Didecyl dimethylammoniumchloride = 8.704% 9.0-10.0 Virex II (256) Johnson/Diversey Foot board of the Lotus rising aid ammonium Alkyldimethylbenzylammoniumchloride = 8.19%...
  • Page 5: Recommendations For Cleaning And Disinfection

    All necessary precautions should be taken to remove all stains and dirt. The cleaning and disinfection method described on the following pages applies specifically to Lopital products and accessories. This method is designed to save time and prevent infections more efficiently.
  • Page 6: Instructions By Product

    Nemo - Shower-toilet Chair 51005100 The Nemo must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
  • Page 7 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Application of common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 8: Flexo - Shower-Toilet Chair

    2017/745 a chair must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
  • Page 9 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Application of common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 10: Elexo - Shower-Toilet Chair

    2017/745 a chair must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
  • Page 11 Disinfecting 5. Repeat the cleaning protocol - Cleaning after reading “Application of common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 6. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 12: Elexo Xxl - Shower-Toilet Chair

    2017/745 a chair must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
  • Page 13 Disinfecting 5. Repeat the cleaning protocol - Cleaning after reading “Application of common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 6. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 14: Reflex - Shower-Toilet Chair

    2017/745 a chair must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
  • Page 15 Disinfecting 5. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 6. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 16: Tango - Shower-Toilet Chair

    Tango – Shower-toilet chair 51005700 The Tango must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
  • Page 17 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 18: Mylo - Shower-Toilet Chair

    Mylo - Shower-toilet chair 51005701 The Mylo must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
  • Page 19 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 20: Tango Xl - Shower-Toilet Chair

    Tango XL – Shower-toilet chair 51005705 The Tango XL must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
  • Page 21 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 22: Tango Xxl - Shower-Toilet Chair

    Tango XXL – Shower-toilet chair 51005800 The Tango XXL must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
  • Page 23 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 24: Timo - Shower(-Toilet) Chair

    Timo – Shower(-toilet) chair 54006020, 54006021, 54006024, 54006030, 54006034, 54006035 The Timo must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products.
  • Page 25 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 26: Cleo - Shower(-Toilet) Chair

    Cleo – Shower(-toilet) chair 54006040, 54006050 The Cleo must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
  • Page 27 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the chair, wipe the chair with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 28: Marina Basic & Marina Deluxe - Shower Trolley

    2017/745 the Marina basic / Marina deluxe must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the trolley thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
  • Page 29 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the trolley, wipe the trolley with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 30: Marina - Shower Trolley

    2017/745 the Marina must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the trolley thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
  • Page 31 Disinfecting 5. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 6. After disinfecting the trolley, wipe the trolley with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 32: Sirocco - Wall-Mounted Shower Trolley

    2017/745 the trolley/dressing table must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the trolley/dressing table thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
  • Page 33 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the trolley/dressing table, wipe the trolley with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 34: Stylex - Shower Trolley

    2017/745 the trolley/dressing table must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the chair thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
  • Page 35 Disinfecting 5. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 6. After disinfecting the trolley/dressing table, wipe the trolley/dressing table with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 36: Luxal - Shower Trolley

    2017/745 the Luxal must be maintained in line with manufacturer’s instructions. This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the trolley thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
  • Page 37 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the trolley, wipe the trolley with a damp cloth and clean water to remove any residue from the disinfectant.
  • Page 38: Lotus - Rising Aid

    Lotus – Rising aid 59008100 The Lotus must be cleaned after every use, with water and regular household cleaning agents. PUR integrated foam components with IMC coating should be rubbed clean and disinfected. Do not use any abrasive or other aggressive cleaning products. If necessary, the PUR components can be disinfected with a solution of 70% ethanol or propanol.
  • Page 39 Disinfecting 4. Repeat the cleaning protocol - Cleaning after reading “Applying common cleaning/disinfectants to Lopital products” on page 3 of this document. Follow the manufacturer's directions for proper use of the disinfectant. 5. After disinfecting the Lotus, wipe the Lotus with a damp cloth and clean water to remove any residue from the disinfectant.