Page 3
5061 JR Oisterwijk T +31 (0)13 5239300 www.lopital.com info@lopital.nl All of the aids produced by Lopital fulfil the following norms for medical products Class 1 (MDR (EU) 2017/745). Explanation of symbols in ‘instructions for use’ and on labels Follow the instructions Note...
1. Introduction 1.1. Introduction The Marina basic / Marina deluxe shower trolley is a medical device for caring for and washing clients in an ergonomically responsible manner. Comfort is the primary consideration for both the client and the caregiver. The Marina basic / Marina deluxe comprises a mobile base frame and a height-adjustable supporting surface with a soft mattress.
· The Marina basic / Marina deluxe is only suitable for moving and caring for one person. · Use only original parts, supplied by Lopital. The use of non-original parts can cause dangers. · The use of components (chargers, cables, actuators, batteries, etc.) that have not been supplied by Lopital may result in increased emissions or decreased immunity of the Marina basic / Marina deluxe (for electric adjustable Marina basic / Marina deluxe).
Page 7
· The Marina basic / Marina deluxe must be serviced and inspected at least once a year. · No modification of the Marina basic / Marina deluxe is allowed, except length adjustment according to the instructions of Lopital. · If the Marina basic / Marina deluxe is adjusted in length, appropriate inspection must be conducted to ensure continued safe use of the shower trolley.
MDR (EU) 2017/745 for medical devices. 1.3. Warranty All Lopital products are supplied with a 2-year warranty, on construction and / or material faults, unless otherwise stated. In the event of varying warranty terms, the exact period will be indicated on your invoice; this also serves as the warranty itself.
3. Operation 3.1 Wheel blocking Marina basic The Marina basic is equipped with 4 swivel wheels. · The brake can be applied on the Marina basic by pushing the blue lip (1) on the wheel downwards (See figure 1). · In order to take the Marina basic off the brake, the same lip on the wheel must move upwards by tapping the upper side of the lip (2) (See figure 2).
· Turn the side rail upwards again to close the side rail (see figure 4). The side rail locks automatically. · The side rail can be placed in 2 positions, wide and narrow. When you fold the side rail upwards, the side rail is first locked in the wide position (see figure 5). By pulling the handle of the lock (see figure 4) you can move the side rail further back to its narrow position.
Page 16
Electrically adjustable versions Moving upwards · If the client is lying on the shower trolley, first put the brake on the wheels and then move the trolley upwards (1) (See figure 8). · Move the trolley upwards by pressing the button that shows an ‘up’...
3.4. Backrest adjustment The Marina deluxe is equipped with a back adjustment. The back adjustment is available with an electrical or mechanical drive. Mechanical back adjustment (hydraulic adjustable Marina deluxe) The back can be moved upwards by means of the handle on the side of the trolley (see Figure 10 Back adjustment figure 10).
3.7. Changing battery (Only electrical adjustable Marina basic / Marina deluxe). The Marina basic / Marina deluxe is an electrically adjustable high-low trolley and works on a battery. The battery is located at the head-end of the trolley. (See figure 14). ·...
1/3 capacity, Fully charged. Charging advised. Charging required, An acoustic signal sounds 2/3 capacity. (red light). 3.8. Charging battery (Only electrical adjustable Marina basic / Marina deluxe). . · Place the battery to be charged on the battery chargestation (see figure 17). ·...
Operation may be continued once the situation is stable again. Turn the emergency stop a quarter turn to unlock it again (See figure 18). First check to ensure the hazard has been removed or the problem has been resolved, before resetting the emergency stop button.
Page 21
Those instruction’s can be found in the maintenance protocol, see website Lopital (www.lopital.com). This process will preferably be carried out by Lopital or another authorised body. Lopital offers the option of having the trolley thoroughly inspected and serviced once a year via Periodic Preventative Maintenance.
Conditions during use 7. Removal of parts All parts that are replaced or removed can be sent back to Lopital or your supplier. You may also offer the parts to the engineer for destruction. We will ensure that all parts are processed in an environmentally-friendly manner.
8. Technical specifications The Marina basic / Marina deluxe is splash-proof; this means that the Marina can be used under the shower. This is given with instruction IPX4. The electronic components used in the Marina basic / Marina deluxe electrical are IPX6. 8.1.
Page 24
Bed base height: minimum 45 cm / 17.7 inch, maximum 95 cm / 23.2 inch Water drainage hose: internal Ø4 cm / 1.6 inch Wheel base: 160 cm / 63 inch Track width: 64 cm / 25.2 inch Undercarriage height: 19 cm / 7.5 inch Marina basic 24 | Instructions for use | Marina basic &...
Page 25
Marina deluxe Marina basic & Marina deluxe | Instructions for use | 25...
8.4. Length adjustment The Marina basic / Marina deluxe is equipped with a length adjustment. For example, the shower trolley can easily be extended or shortened. If you want to adjust the length of your Marina basic / Marina deluxe, we recommend that you have this done by your sup- plier.
- Drainage hose - Lying plates If other parts needs to be replaced, this must be done by authorised personnel. Contact your dealer or contact Lopital. Mattress 1. 1. To remove the mattress you have to loosen the push Figure 22 Replacement Mattress.
90 degrees to the right. Marina deluxe stops. Technical issue. Contact Lopital or your supplier. For all other issues please contact Lopital or your supplier.: Telephone: +31 (0)13 5239300 E-mail: info@lopital.nl 28 | Instructions for use | Marina basic & Marina deluxe...
10. Relation with other supports Slide sheets A slide sheet is the ideal device for transferring from bed to trolley and vice versa. The physical burden and time required when using a slide sheet is minimal. A slide sheet is suitable for most bed and chair transfers from and to the Marina basic / Marina deluxe.
11. Electromagnetic compatibility Professional healthcare environment Guidance and Manufacturer’s declaration-electromagnetic emissions The Marina basic / Marina deluxe is intended for use in the electromagnetic environment specified below. The customer or the user of Marina basic / Marina deluxe should assure that it is used in the following environment: Emissions Test Compliance Electromagnetic Environment - guidance...
Page 31
Guidance and Manufacturer’s declaration-electromagnetic immunity The Marina basic / Marina deluxe is intended for use in the electromagnetic environment specified below. The customer or the user of the Marina basic / Marina deluxe should assure that it is used in the following environment: Immunity Test IEC 60601-1-2 Test Compliance Level...
Page 32
32 | Instructions for use | Marina basic & Marina deluxe...
Page 33
Marina basic & Marina deluxe | Instructions for use | 33...
Page 34
For more information, visit our website: www.lopital.com Lopital Nederland B.V. Laarakkerweg 9, 5061 JR Oisterwijk, Postbus 56, 5060 AB Oisterwijk Tel +31 (0)13 5239300, E-mail info@lopital.nl, www.lopital.nl User2000-02-2022...