7.2 Battery Status Indicator

Battery status will be displayed in section 1 of the display (Figure 4). There are
two different status indicators showing battery discharging (Table-7.1) and
charging (Table-7.2) status respectively.
Table-7.1 Battery discharging status
1
2
3
4
5
Table-7.2 Battery charging status
1
2
3
4
5
Warning
At approximately 2% of full battery capacity, the ventilator will not start
when it is turned off or will shut down when it is turned on.
Note
Battery capacity levels are detected from measured voltages and the
capacities shown above are based on results from new batteries tested at
room and low temperature. Battery capacity levels are subject to change when
old batteries are used.
Fully charged Batteries must be recharged after 6 months without usage or
when battery discharging LED turns orange and flashing. Battery Re-charging
time is about 5.5 hours from fully discharged. Batteries have a minimum of 200
discharge and charge cycles.
Full Capacity
Approx. 75% of full
capacity
Approx. 50% of
full capacity
Approx. 25% of full
On screen symbol change to Yellow color and
capacity
Approx. 5% of full
On screen symbol change to Red color and
capacity
flashing with associated Red color warning LED.
Full Capacity
95% of full
capacity
90% of full
capacity
80% of full
capacity
65% of full
capacity
e600 User Manual
No Alarm
No Alarm
No Alarm
flashing
No Alarm
No Alarm
No Alarm
No Alarm
No Alarm
(15PL1010 - Rev.7 28_8_2017)