Trouble Shooting - O-Two e600 User Manual

Automatic transport ventilator

Advertisement

Chapter 10
Warning
Please contact manufacturer if a problem cannot be rectified. For the safety of
the patient and the Health care providers DO NOT continue using the
ventilator.
Message/fault
Leaking mask, ETT,
disconnected patient
circuit
Low Airway Pressure
High Airway Pressure
Apnea, spontaneous
breathing failed or
disconnection, faulty
sensor
No Oxygen ≤ 20 PSI
Low Oxygen(40-21
PSI)
Leakage, measured
expiratory volume is
40% lower than set.
Invalid setting
Battery
No proper
discharges
charging/Faulty
quickly
battery
Battery empty/no
e600 cannot
power supply
be switched on
connected/defective

Trouble shooting

cause
e600 User Manual
Remedy
Make sure mask fits / all patient circuit
parts are properly connected./change
patient circuit
Replace hose, check tube position,
check ventilation settings and correct
Check patient, ventilation circuit,
check tube position, adjust Pmax
alarm value
Switch to Assist Control ventilation,
ensure connections tight, Replace
patient circuit
Change Oxygen cylinder
Change Oxygen cylinder
Check leakage in breathing system.
Resolve parameters conflict by
resetting out of range parameters
Charge Battery as per
instructions/Replace battery
Change Battery/connect power
supply/ send to O-Two for repair or
service
(15PL1010 - Rev.7 28_8_2017)
47
English version