5.2.2 SIMV (Synchronized Intermittent Mandatory Ventilation).
In SIMV mode the ventilator will deliver volume ventilation at the set
Tidal Volume (V
t
The default trigger for this mode is 3 L/min but can be adjusted up to 15
L/min. if trigger condition is met, the ventilator will deliver synchronized
volume controlled mandatory ventilation.
In SIMV mode the selected breathing rate remains constant and the time
of spontaneous breathing window will change if patient triggers the
synchronized mandatory breath before the normal start of inhalation
phase (beginning of Ti).
If no effort was detected during the trigger period, the ventilator will
initiate mandatory ventilation at the end of trigger window (Figure 11).
Should the patient demand more flow than set by user, he/she can
withdraw the excess from ambient.
Table-3 Default Ventilation Setting- SIMV
Parameter
Tidal Volume
(50 - 2000 ml)
Rate
(5 - 60 BPM)
Mv
Calculated based on Vt & f
I:E ratio
Ti*
PEEP
(OFF, 4-20 cm H
Trig.
(1 -15 L/min)
O
%
(100% or 60%)
2
P max.
10 -80 cm H
P min.
0-20 cm H
Refer to Manual and I-Hold
Manual
* Ti may be limited below its range depending on set I:E ratio and rate.
) and Rate (BPM).
Range
(1:4 – 3:1)
(0.2 - 9 sec)
O)
2
O
2
O
(during I time only)
2
section
e600 User Manual
Default
Infant
Child
100
250
30
15
3.0
3.7
1:2
1:2
0.66
1.33
5
5
3
3
100
100
25
25
3
3
ready
ready
(15PL1010 - Rev.7 28_8_2017)
Adult
500
10
5.0
1:2
2.0
5
3
100
30
3
ready
24
English version