Note
 The Battery pack can be charged during operation.
 The Battery pack may be charged using external power supply
supplied with the unit or the optional DC to DC power supply.
 The Battery pack should be replaced after 200 charge/discharge
cycles or if the battery pack will not fully charge (as indicated on the
battery display on the ventilator) or if the ventilator doesn't run for
more than 5 hours on a single charge.
8.3

Ambient Air Entrainment Filter

The e600 entrains ambient air through the internal Venturi system for
ventilation when the O
decreased oxygen concentration but also increases the ventilator operating
time on an oxygen cylinder.
Caution
Always keep the ambient air entrainment port clear of obstructions. Always
replace the filter after use.
Warning
Avoid particulate and/or gaseous pollutants in the ambient air! The
entrainment of pollutants into the ventilator may cause the ventilator to
malfunction or cause danger to the patient.
8.4 Preventive Maintenance and Servicing
It is recommended that routine preventive maintenance (PM) and
servicing shall be carried out as per the following:
Type
Description Procedure
Charging
PM
battery
PM
Leak test
Function
PM
check
Level II

Servicing

service
Servicing Full service
concentration is set at 60%. This provides not only
2
Criteria
User Manual
Battery fully
Chapter 8.2
charged
User Manual
No leak observed
Chapter 4.2
User Manual
No abnormal
Chapter 4.2
function observed
Service
Meet product
manual
specifications
Service
Meet product
manual
specifications
e600 User Manual
Schedule
Every 6
months
Every 6
months
Every 6
months
Manufacturer
Every 2
or authorized
years
service center
Every 6
Manufacturer
years
(15PL1010 - Rev.7 28_8_2017)
By
User
User
User
38
English version