O-Two e600 User Manual page 27

Automatic transport ventilator

Advertisement

Table-2 Default Ventilation Setting- A/C V
Parameter
Tidal Volume
Rate
Mv
Calculated based on Vt & f
I:E ratio
Ti*
PEEP
Trig.
O
%
2
P max.
P min.
0-20 cm H
Manual
Refer to Manual and I-Hold section
* Ti may be limited below its range depending on set I:E ratio and rate.
Range
(50 - 2000 ml)
(5 - 60 BPM)
(1:4 – 3:1)
(0.2 – 9 sec.)
(OFF, 4-20 cm H
O)
2
(OFF,
1
-15 L/min)
(100% or 60% O2)
10 -80 cm H
O
2
O during I time only
2
Figure-10 A/C V waveform
e600 User Manual
Default
Infant
Child
100
250
30
15
3.0
3.7
1:2
1:2
0.66
1.33
5
5
3
3
100
100
25
25
3
3
ready
ready
(15PL1010 - Rev.7 28_8_2017)
Adult
500
10
5.0
1:2
2.0
5
3
100
30
3
ready
23
English version