Electromagnetic Immunity
Immunity test
Electrostatic discharge
(ESD) IEC61000-4-2
Electrical fast transient/
burst
IEC61000-4-4
Surge IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
Power frequency (50/60
Hz) magnetic field
IEC61000-4-8
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
O-Two e600 is intended for use in the electromagnetic environment in which
radiated RF disturbances are controlled. The user can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the O-
Two e600 as recommended below, according to the maximum output power
of the communications equipment:
Rated
maximum
150kHz to 800
MHz outside
output power
ISM bands d =
of transmitter
3.5/3 * √ P
W
0.01
0.1
1
10
100
IEC60601 required
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV line to line;
± 2 kV line to earth
<5%U
for 0.5 cycle
T
40%U
for 5 cycle
T
70%U
for 25 cycle
T
<5%U
for 5 s
T
3 A/m
3Vrms: 50kHz to 80
MHz
outside
bands
10Vrms: 150kHz to 80
MHz in ISM bands
10 V/m @
80 MHz to 2.5 GHz
Separation distance (m)
150kHz to 800
MHz in ISM
bands d = 1.2 √ P
0.12
0.12
0.4
0.4
1.16
1.2
3.8
3.8
12
12
e600 User Manual
Actual compliance
level
± 8 kV contact
± 15 kV air
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 1 kV line to line;
± 2 kV line to earth
<5%U
for 0.5 cycle
T
40%U
for 5 cycle
T
70%U
for 25 cycle
T
<5%U
for 5 s
T
30 A/m
3 Vrms
ISM
10 Vrms
30 V/m
80
MHz
to
800
800 MHz
2.5 GHz
d = 1.2 √ P
d = 2.3 √ P
0.12
0.4
1.2
3.8
12
(15PL1010 - Rev.7 28_8_2017)
MHz
to
0.23
0.7
2.3
7.3
23
45