Download Print this page

O-Two e600 User Manual page 25

Automatic transport ventilator

Advertisement

AC/V Settings
Rate (BPM)
Vt (ml)
Mv (L)
(calculated values)
PEEP (cm H
O)
2
P max (cm H
O)
2
P min (cm H
O)
2
Ti (Sec.)
I:E
O
(%)
2
F trig. (L/min)
A/C V mode default settings
Table-1
A/C V mode screen w/ default parameters (Child setting)
Infant
Child
30
15
100
250
3.0
3.7
5
25
25
3
0.66
1.33
1:2
1:2
100%
100%
3
Figure-7
Figure-8
e600 User Manual
(15PL1010 - Rev.7 28_8_2017)
Adult
10
500
5.0
5
5
30
3
3
2.0
1:2
100%
3
3
Showing frame
around
Parameter during
selection
21
English version

Advertisement

loading

This manual is also suitable for:

E series