Automatic transport ventilator

Advertisement

Cautions
 When the ventilator is not in use, always turn off the gas supply.
 Never allow oil or grease to come into contact with any part of the
cylinder, regulator, or ventilator.
 After use, always ensure that all components are cleaned in accordance
with the instructions provided in this manual. (See section 8.1 - Cleaning
and Disinfection).
 The use of this device in a carrying case may result in an increase in
oxygen concentration or lower than intended ventilation volume when
used in the 60 % air mix mode. When air mix mode is being used it is
recommended that the ventilator be placed in its normal operating
orientation and the air intake on the side of the ventilator is not
obstructed.
 Never operate the ventilator without an intake filter being fitted,
otherwise particles may contaminate the ventilator and affect its
function.
 This ventilator must only be serviced by the manufacturer or its
authorized service centers.
 
 Device, single use patient circuits and battery packs should be safely
discarded in accordance with local state and institutional laws and
procedures. .
 
 Spare O-Two patient circuits are recommended at all times.
 
 The design of this Ventilator does not incorporate a negative pressure
phase during either automatic or manual ventilation.
 
 The external power supply and battery are component parts of the
medical electrical equipment system.
 
 The ventilator is considered as a high flow device as its maximum flow
output at pressure of 40.6 PSI is around 75 L/min. It should only be
connected to a pipeline system that allows for the indicated high flow to
avoid interfering with the operation of adjacent equipment.
 
 Always ensure that all components are assembled correctly and ready
for use.
 
 Exhaled volume of the patient can differ from the measured exhaled
volume due to leaks around the mask.
 When selecting very small tidal volumes during ventilation of infants
with, take into consideration the dead space in the patient circuit.