Functional Check - LeCroy ZS1000 Instruction Manual

High impedance active probe
Table of Contents

Advertisement

6. Disconnect the ProBus Extender Cable from the oscilloscope. Verify
that the scale factor changes from 20 mV/div to 2 mV/div.
7. Reconnect the ProBus extender Cable to the oscilloscope.
The warranted characteristics of the ZS1000 are valid at any tempera-
ture within the Environmental Characteristics listed in the Specifications.
However, some of the other test equipment used to verify the perfor-
mance may have environmental limitations required to meet the accu-
racy needed for the procedure. Make sure that the ambient conditions
meet the requirements of all the test instruments used in his procedure.
The correct operation of the ZS1000 controls requires software version
5.0.0.2 or higher. The software version in the test oscilloscope can be
verified by selecting Utilities, Utilities Setup... from the menu bar, then
the Status tab.
Contact your local LeCroy representative or visit www.lecroy.com if the
software in your oscilloscope requires updating.

FUNCTIONAL CHECK

The functional check will verify the basic operation of the probe func-
tions.
It is recommended that the Functional Check be performed prior to the
Performance Verification Procedure.
1. Return to the factory default settings by:
a. Selecting File, Recall Setup... from the menu bar.
b. Then touching the Recall Default button.
3. Touch the C1 trace label to open the C1 Vertical Adjust dialog.
4. Verify that the probe sensed (ZS1000) is displayed as a dialog tab.
ZS1000-OM-E Rev C
Performance Verification
Note
7–3