Menu Item Descriptions

Display
1. Measurement - Show the primary ppb
2. Diagnostics - (Service passcode required)
Output
1. Rcd Zero - Specify the ppb reading equivalent to 4 mA
2. Rcd Span - Specify the ppb reading equivalent to 20 mA
3. Rcd Trim - Adjust/calibrate the 4-20 mA output
a. Reset Trim - Clear trim values prior to recalibrating output
b. Trim Zero - Adjust the 4 mA output
c. Trim Span - Adjust the 20 mA output
4. Test - Force the 4-20 output to a specified percent of scale
User - (User passcode required)
1. PPB Offset - Enter PPB Offset
2. Clear PPB Offset - Clear PPB Offset
3. Sensor Dry Down - Dry the Sensor
Comms
1. Baud Rate - Specify the data rate for RS485 communications
2. Node ID - Specify the PanaView Network Node ID
3. Test - Force meter to transmit an ID message for test purposes
Service - (Service passcode required)
1. Update - Update the unit software.
2. Params - Access the parameter storage.
a. Dump - Dump the parameter storage.
b. Erase - Erase the EPROM.
3. Heater Test - Test the heater.
a. 50°C to 190°C - Select the temperature for testing.
b. OFF - Turn the heater off.
c. Resume - Resume normal measurement operation.