General Precautions During
Patient Care
WARNING !
Do not cover phototherapy lights or incubator
hood with blankets, aluminum foil, or other
materials intended to boost the effect of
phototherapy. This would cause build-up of
heat since the incubator would not be cooled
by ambient air, a factor absolutely essential
for operation. - Danger of overheating patient!
Preventing high noise levels
Excessive noise for the patient may be caused by:
– use of head boxes and/or delivery of pressurized
gases,
– objects dropped on the incubator hood.
Therefore:
Do not place anything on the incubator hood.
Electrical safety
WARNING !
This device is to be used only in rooms with
line power installations complying with
national safety standards for hospital patient
rooms (e.g. IEC 601.1 "Safety of Medical
Equipment).
To maintain grounding integrity, connect only
to a "hospital grade" receptacle.
To reduce risk of electric shock, do not
remove cover.
Always disconnect supply before servicing.
Use only electromedical accessories complying with
CSA 601-1, IEC 601-1, EN 60601-1 or DIN VDE 0750,
part 1, UL 544 or other local or state regulations
governing the use of electromedical equipment in
hospitals.
Operation
35