Dräger Incubator 8000 NC Operating Instructions Manual page 38

Hide thumbs Also See for Incubator 8000 NC:
Table of Contents

Advertisement

Operation
Extending Temperature Range From 37 ºC to 39 ºC
WARNING !
Temperatures above 37 °C may only be used
on the order of a physician.
In this mode of operation, the baby´s tempera-
ture must be monitored especially carefully.
1
Press unlocking button
yellow indicator LED >37 °C lights up.
2
NOTE: This alters the temperature threshold for heater
cut-off in the event of overtemperature from 38 °C to
40 °C.
P
p
3
Keep
or
key pressed until the value desired is
displayed.
NOTE: Set value between 37.1 °C and 39 °C within the
next minute.
If extended range settings are not used within one minute
– control LED >37 °C will go off,
– blocking of extended temperature range will be
reactivated,
– the threshold for overtemperature will be reset to
38 °C.
Using a set value below 37 °C after use of the extended
range will also automatically cancel range exension.
NOTE: In case the incubator temperature is still above
38 °C when leaving the extended range, the overtem-
perature alarm (see page 39) would be triggered.
To avoid this alarm: First set value to 37.1 °C and allow
incubator to cool below 38 °C. Then set desired value
below 37 °C.
38
7
,
Using Air Temperature Control
°C LUFT/AIR
2
°C
1
>37 °C
3
3
Check
Reset
Control