Turn Capstan Fully Clockwise - Fluke Calibration P3800 Series User Manual

High pressure hydraulic deadweight tester
Table of Contents

Advertisement

1. Open HP valve (1) and LP valve (2) fully, (counter-clockwise).
2. Open reservoir valve (3), (approximately 4 turns counter-clockwise).
3. Turn capstan fully out (counter-clockwise), and wait for approximately 60 seconds.
4. Turn capstan fully in (clockwise), and wait for approximately 15 seconds.
5. Trapped air from the system will appear as bubbles in the reservoir.
6. Repeat the procedure until no more air appears in the reservoir.
information@ITM.com
Figure 3-4. Turn Capstan Fully Clockwise
Fluke-Direct
Procedure for Removing Trapped Air from System
.ca
3
Operation
glf05.bmp