Spectrum Analyzer Specifications (Option 102) - HP 8920A Application Handbook

Rf communications test set
Hide thumbs Also See for 8920A:
Table of Contents

Advertisement

Spectrum Analyzer Specifications (Option 102)

Spectrum Analyzer Specifications (Option 102)
Frequency
Table 4
272
Frequency Range: 400 kHz to 1 GHz
Frequency Span/Resolution Bandwidth (coupled):
Span
Bandwidth
< 50 kHz
300 Hz
< 200 kHz
1 kHz
< 1.5 MHz
3 kHz
< 18 MHz
30 kHz
≥ 18 MHz
300 kHz
Full span
Display: Log with 10 dB/div, 2 dB/div, or 1 dB/div
Display Range: 80 dB
Reference Level Range: +50 to −50 dBm
Residual Responses: <−70 dBm (no input signal, 0 dB attenuation)
Image Rejection: >50 dB
Non-harmonic Spurious Responses: >70 dB (for input signals ≤−30 dBm)
Level Accuracy: ±2.5 dB
Displayed Average Noise Level: <−114 dBm for ≤ 50 kHz spans
Log Scale Linearity: ±2 dB (for input levels ≤ −30 dBm or 60 dB range)