HP 8920A Application Handbook page 148

Rf communications test set
Hide thumbs Also See for 8920A:
Table of Contents

Advertisement

AM Receivers
148
On the Test Set using the knob and data entry keys:
15. Slowly increase the Amplitude until squelch just remains open.
Critical Squelch is displayed as Amplitude as shown.
Record this level.
Squelch
Sensitivity
16. Decrease Amplitude until the Receiver's squelch just closes.
Record this level.
Calculate the measurement:
17. Critical Squelch Hysteresis is the difference between the two readings.
On the Radio set the Receiver's Controls as follows:
18. Set the Receiver's squelch control to maximum.
On the Test Set using the knob and data entry keys:
19. Increase Amplitude until the Receiver's squelch just opens.
Record this level.
20. Record the maximum squelch sensitivity is displayed as Amplitude as
shown.
21. Decrease Amplitude until the Receiver's squelch just closes.
Record this level.
Calculate the reading:
22. Maximum Squelch Hysteresis is the difference between the two readings.
RX TEST
SINAD
dB
0
20.44
24
RF Gen Freq
AFGen1 Freq
AFGen2 Freq
28.310000
1.0000
1.0000
MHz
kHz
AFGen1 To
AFGen2 To
Amplitude
1.58
AM
FM
uV
30.0
Off
%
Atten Hold
On/Off
Output Port
RF Out/Dupl
AC Level
V
0.7878
To Screen
Filter 1
RF GEN
50Hz HPF
kHz
RF ANL
AF ANL
Filter 2
SCOPE
15kHz HPF
SPEC ANL
ENCODER
DECODER
Ext Load R
RADIO INT
8.00
More

Advertisement

Table of Contents
loading

This manual is also suitable for:

8920b

Table of Contents