HP 8920A Application Handbook page 158

Rf communications test set
Hide thumbs Also See for 8920A:
Table of Contents

Advertisement

AM Receivers
158
On the Signal Generator:
17. Set RF Power to ON.
18. Set Output Level to 31.6 mV.
19. Set Output Frequency to lowest frequency being checked (typically
<1 MHz).
20. Set Modulation to internal AM.
21. Set Modulation rate to 400 Hz.
22. Set Modulation Depth to 30%.
23. Tune the Signal Generator in 1 kHz steps from the minimum to maximum
frequency and note frequencies where SINAD decreases below 10 dB.
If necessary, increase tuning resolution on the Signal Generator and ad-
just as required to locate frequency causing maximum degradation of
the SINAD reading.
24. For each frequency where SINAD is degraded, reduce the Signal Generator
RF output level until SINAD of 10 dB is obtained.
Record the Signal Generator frequency and level.
25. Repeat
step
23.and
Calculate the Spurious Response:
26. The minimum result is the calculated spurious response attenuation.
Each Spurious response level recorded in step 25
calculated spurious response attenuation
step 24.
until all spurious responses are measured.
Amplitude level recorded step 17