Chapter 8, Programming The Call Processing Subsystem
Programming The ANALOG MEAS Screen
AFGen1 To
Amplitude
De-Emphasis
Detector
Filter 1
534
This field has two subfields:
•
the upper subfield sets the destination port for Audio Frequency Generator #1
•
FM = RF Generator FM modulator
•
AM = RF Generator AM modulator
•
Audio Out = AUDIO OUT connector on front panel of Test Set
•
the lower subfield sets the:
•
FM modulation deviation if upper subfield set to FM
•
AM modulation depth if upper subfield set to AM
•
amplitude of audio signal (volts RMS) at the AUDIO OUT connector if upper sub-
field set to Audio Out
For testing mobile stations the upper field is normally set to
set to the desired FM deviation in kHz.
for programming command syntax.
104
This field sets the output power of the Test Sets's transmitter (that is, the output
power of the Test Set's RF Generator).
for programming command syntax.
167
This field is used to select or bypass the 750 uSec de-emphasis filter network used
to condition the audio signal before being analyzed by the Audio Frequency
Analyzer.
See "AF Analyzer," in chapter , on page 101
syntax.
This field is used to select the type of detector used to measure the amplitude of
the audio signal being analyzed by the Audio Frequency Analyzer.
Analyzer," in chapter , on page 101
This field selects one of several standard or optional audio frequency filters which
can be used to condition the audio signal before being analyzed by the Audio
Frequency Analyzer.
See "AF Analyzer," in chapter , on page 101
command syntax.
See "AF Generator 1," in chapter , on page
See "RF Generator," in chapter , on page
for programming command
for programming command syntax.
and the lower field
FM
See "AF
for programming