HP 8921A Programmer's Manual page 278

Cell site test set
Hide thumbs Also See for 8921A:
Table of Contents

Advertisement

Chapter 5, Advanced Operations
Status Reporting
Calibration Status Register Group
:STATus: CALibration: CONDition ?
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Condition
Transition
Filter Registers
Register
Figure 14
Calibration Status Register Group
278
The Calibration Status Register Group contains information about the Test Set's
hardware. This status group is accessed using the STATus commands. The
Calibration Status Register Group uses 16-bit registers and includes a Condition
Register, Transition Filters, an Event Register, an Enable Register, and a
Summary Message. Refer to the
for a discussion of status register operation.
244
STATus commands for the Calibration Status Register Group.
:STATus: CALibration
:STATus: CALibration: EVENt ?
Event
Register
"Status Reporting Structure Overview" on page
?
:PTRansition
<integer>
?
:NTRansition
<integer>
:STATus: CALibration: ENABle
Enable
Register
shows the structure and
Figure 14
?
<integer>
Summary Message Bit
(to bit 8 of Questionable
Data/Signal Register Group
Condition Register
ch4drw11.drw

Hide quick links:

Advertisement

Table of Contents
loading