Fluke 150X Series Calibration Manual page 27

Insulation testers
Table of Contents

Advertisement

150X Series
Calibration Manual
Cal Keypad
Keys Used in the Calibration Steps
The Tester keys behave as follows when you do the calibration adjustment procedure:
Tstores the calibration value and advances to the next step.
H ignores the calibration value and advances to the next step. H is also used to exit the
calibration function after the calibration adjustment sequence is complete.
Calibration Adjustment Procedure
Use these steps to make calibration adjustments to the Tester. If the Tester is turned off
before completion of the adjustment procedure, the calibration constants are not changed.
1. Remove batteries from DUT battery compartment. Connect a +5.0 V lab supply to the + and
− battery terminals.
2. Hold down L, turn the rotary function switch from OFF to VAC. The Tester displays CAL.
3. Press the blue button once to see the calibration counter.
4. Press blue button again to start the password entry. --- appears on the display.
5. Press four keys to enter the password.
6. Press the blue button to go to the first calibration step. The Tester displays C-01 if the
password is correct. If the password is not correct, the Tester emits a double beep,
displays ---- and you must enter the password again. Repeat step 4.
22
Figure 6. Restore the Default Password
Battery Pads