Verathon BLADDERSCAN BVI 3000 Service Manual page 23

Hide thumbs Also See for BLADDERSCAN BVI 3000:
Table of Contents

Advertisement

SCAN THE BLAddER PHANTOM
40. Ensure there is sufficient water on the scan membrane of the adult bladder phantom. The entire
membrane should be covered.
41. On the console, on the Main Configuration screen, press the RUN button. The console is now in
standard operating mode.
THIS UNIT IS NOT SETUP
CABLE
TEST
42. On the Main Menu screen, press the SCAN button.
05/22/14
SCAN
43. On the Scan screen, press the Male/Female button in order to select male mode.
READY TO SCAN
DONE
44. Place the dome section of the probe onto the phantom scan membrane, and then press it gently against
the top surface of the phantom.
45. On the probe, press the Scan button. The probe begins to scan the phantom.
46. Hold the probe steady throughout the scan. The console clicks once at each of the twelve scan planes.
When the console beeps, the scan is complete, and the Aiming screen is displayed.
132ml
READY TO SCAN
CURRENT = 132ml
DONE
47. On the console screen, verify that the measurement is accurate to ± (10% + 10 mL) of the phantom's
volume.
48. On the certificate, make note of the scan result.
49. Repeat Step 42 through Step 48 in female mode, and then confirm the female mode measurement is
within the accuracy range.
Verathon Service Partner Manual: Annual Certification
PROG
MFG
SETUP
PROBE
TEST
NOW
BVI 3000
VERATHON
ALT
PRINT
SCAN
SCAN
X
RUN
13:27
NOTES
15