P103 Cancel Mounting Position Effect; P104 Adjustment Pressure Unit On Display - KLAY-INSTRUMENTS 2000-VALVE-RANGE Series Instruction Manual

Pressure-and level transmitter
Table of Contents

Advertisement

Instruction Manual Series 2000-RANGE-VALVE
P103
CANCEL MOUNTING POSITION EFFECT
All transmitters are calibrated vertically. When a transmitter of the Series 2000-VALVE-
RANGE is installed horizontally, there will be a small "mounting effect" on the zero
(4 mA). For example the transmitter shows 4.03 mA instead of 4.00 mA.
This can be easily canceled with P103. In P103 there are three options:
1.
2.
3.
When "SET" is selected the transmitter is automatically adjusted at 4.00 mA. The
span will not be affected.
CAUTION: Do not apply pressure while executing "cancel mounting position effect"
N.B.:
From revision 2 the description above is valid. Transmitters working with software
revision 1 will direct process cancel mounting position.
P104
ADJUSTMENT PRESSURE UNIT ON DISPLAY (See Conversion table)
Several engineering units can be shown on the display by using a conversion factor.
(See conversion table below). As standard the pressure unit of the transmitter is set to
bar. Four engineering units can be used for displaying the applied pressure on the
display (mH
2
1. Press [PROG] until "100" appears on the display.
2. Push 4x
3. Now press [PROG] to confirm this.
4. Now press
table below. The number in brackets, see Display column, represents the menu
options of transmitters with other software. Example: 3 = bar.
5. Then press [PROG] again to confirm. The transmitter is now set to read in "bar".
Conversion table:
DISPLAY:
mH2O (mWK) * (1)
mmH2O (mm WK) (2)
bar * (3)
Mbar (4)
PSI * (5)
Atm (6)
KPa (7)
MPa (8)
Kgf/cm2 (9)
mm HG (10)
H/EN/2000-VALVE/02-2021/05
ESC:
Nothing can be changed.
Leave without doing anything. (confirm with PROG).
RESET:
Use this option when you are not sure if P103/SET has been done in a proper way.
(confirm with PROG).
when using this option the original factory setting will be valid.
SET:
Canceling mounting position effect. (confirm with PROG).
O, inH
O, bar and PSI).
2
[]
to get to point P104 (pressure unit setting).
[]
[]
or
and set it to the correct pressure unit. See also the conversion
Klay Instruments B.V.
CONVERSIE FACTOR:
1.000
1000
0.09806
98.0665
1.4223
0.0967
9.80665
0.009807
0.1
73.556
Page 11