Siemens MOBILETT Plus HP Installation And Setting Instructions Manual page 9

Table of Contents

Advertisement

General
Fig. 3
Disconnect batteries by removing one of the KBATT plugs (refer Fig. 3 and Fig. 4).
Measure the residual voltage at test points CAPPOS and CAPNEG on board D7 (refer to
Fig. 2). The voltage measured must be less than 2 VDC.
WARNING
- Make sure that the correct measurement range is set on the measurement device.
- Connect the measurement device to CAPPOS and CAPNEG.
- Switch S10 off. Connect the mains cable and switch the system ON.
- Check whether the voltage increases.
- Switch the system OFF, disconnect the cable and switch S10 on.
- Check whether the voltage decreases.
Siemens-Elema AB
Solna, Sweden
If the voltage measured between CAPPOS and CAPNEG is 0V, the
measurement device could be defective or the wires between the
CAPPOS and CAPNEG test points on the capacitor bank could be
damaged.
The capacitors could still be charged.
In this case, proceed as follows:
SPR8-220.033.03
Rev. 02
Fig. 4
Page 5 of 8
10.02
SPS-UD
1 - 5
MOBILETT Plus HP

Advertisement

Table of Contents
loading

Table of Contents