Installation and setting
Fig. 1
Unpacking, applying warning labels
WARNING
Fill in the IVK list or report the IVK components to your responsible uptime service center
(USC).
Unpack the protractor and place it on a proper place e.g. the column.
Connect the exposure release cable.
Mount warning label 60 77 411. Choose appropriate language version and paste it/them
on the Control and display panel according to X, Y / Fig. 1.
Only in Germany:
Mount label 96 33 645 on the lower arm 50 mm above the link, see Fig. 2.
NOTE
Siemens-Elema AB
Solna, Sweden
Important when unpacking:
Risk for serious damages due to humidity condensation or to
high temperature inside the batteries.
Assure that the unit has established normal room temperature.
The single tank temperature must be >15°C before startup.
If, due to increased operating loads (DIN 6815), there are other
dimensions on site for the control range, appropriate labels can
be ordered:
- Control range: 2.5 m, Part No. 84 14 823
- Control range: 3.5 m, Part No. 31 36 731
SPR8-220.033.03
Rev. 02
Fig. 2
Page 1 of 12
10.02
SPS-UD
2 - 1
2
50 mm
MOBILETT Plus HP