Pressure Channel Is Not Displayed; Pressure Measurement Is Not Accurate; One Pressure Channel Measures Incorrectly - LABORIE Solar Blue Service & Installation Manual

Table of Contents

Advertisement

4.5.5

Pressure channel is not displayed

One of the pressure channels (e.g. Pves, Pabd, Pura, Pdet,) is not displayed
in the measurement program.
 Select the channel via the protocol settings and channel settings.
4.5.6

Pressure measurement is not accurate

The pressure recording is not accurate.
 Is the pressure transducer (solid state catheter, water transducer, air-
charged transducer) connected to the Solar Blue module?
 Are the pressure transducers connected to the right channels (P1=Pves,
P2= Pabd, P3=Pura)?
 Is the setup for measuring with water catheters correct?
 Check if one of the cables, plugs or connectors is damaged.
 Check if one or more transducers are defective, using the pressure
transducer test.
 Contact your LABORIE representative.
4.5.7

One pressure channel measures incorrectly

One pressure channel measures incorrectly during an investigation.
 Most probably the pressure channel was not properly zeroed.
 Connect another Lemo 4 connector of a solid-state catheter. Check if
something is wrong with the catheter or the Solar Blue interface.
 Connect the air charged catheter to another pressure transducer. Check if
something is wrong with the transducer/catheter or the Solar Blue interface.
 Connect the water catheter to another pressure transducer. Check if
something is wrong with the transducer/catheter or the Solar Blue interface.
70
© 2008 – 2020 by MMS B.V.