4.5.5
Pressure channel is not displayed
One of the pressure channels (e.g. Pves, Pabd, Pura, Pdet,) is not displayed
in the measurement program.
Select the channel via the protocol settings and channel settings.
4.5.6
Pressure measurement is not accurate
The pressure recording is not accurate.
Is the pressure transducer (solid state catheter, water transducer, air-
charged transducer) connected to the Solar Blue module?
Are the pressure transducers connected to the right channels (P1=Pves,
P2= Pabd, P3=Pura)?
Is the setup for measuring with water catheters correct?
Check if one of the cables, plugs or connectors is damaged.
Check if one or more transducers are defective, using the pressure
transducer test.
Contact your LABORIE representative.
4.5.7
One pressure channel measures incorrectly
One pressure channel measures incorrectly during an investigation.
Most probably the pressure channel was not properly zeroed.
Connect another Lemo 4 connector of a solid-state catheter. Check if
something is wrong with the catheter or the Solar Blue interface.
Connect the air charged catheter to another pressure transducer. Check if
something is wrong with the transducer/catheter or the Solar Blue interface.
Connect the water catheter to another pressure transducer. Check if
something is wrong with the transducer/catheter or the Solar Blue interface.