Roam Nxt, Pim Nxt For Fluid, And Transducer Setup; Figure 13: Roam Nxt, Pim Nxt For Fluid, And Pressure Transducer Connected - LABORIE NXT Pro Owner's Manual

Hide thumbs Also See for NXT Pro:
Table of Contents

Advertisement

IMPORTANT: T-DOC Air-Charged Catheters are intended for single-use. Synergy will
not recognize a catheter that has been previously registered in the software after the
allotted two hours to complete urodynamic testing.
Confirm the preferred catheter placement. Then, connect the infusion pump tubing
to the filling lumen on the vesical catheter or filling catheter. Ensure that all air
bubbles have been removed from the infusion pump tubing and that the infusion
pump has been properly primed before connecting the line to the filling lumen on
the vesical catheter or filling catheter.
For information on T-DOC catheters, refer to the section
page 115.
2.6.2.2

Roam NXT, PIM NXT for Fluid, and Transducer Setup

Ensure that the PIM NXT for Fluid is connected to the Roam NXT. A solid-green
light bar on the Roam NXT indicates that the device is ready for use. For additional
information on connecting the PIM NXT to the Roam NXT, refer to the section
Replacing the PIM NXT
Use the PIM NXT for Fluid to Medex Transducer Cables to connect the pressure
transducers to the Pressure Channel Connectors on the PIM NXT for Fluid, as
shown in Figure 13.

Figure 13: Roam NXT, PIM NXT for Fluid, and Pressure Transducer Connected

IMPORTANT: The Pressure Channel Connectors on the PIM NXT for Fluid are color
coded blue for Pves, yellow for Pura, and red for Pabd. The green color-coded channel is
capable of Pves, Pura, or Pabd measurements.
Use Medex clamps and brackets to mount the pressure transducers on the NXT
hoop or an IV pole at the patient's waist height or higher.
Remove the disposable cartridges from their packages. Remove the protective
covers from the back and from the connector ends of the cartridges. Slide a
cartridge over each transducer until it clicks into place (Figure 14).
Aseptically remove pressure measurement tubing from its package and attach the
three-way stopcock to the lower (male) end of the pressure cartridge.
To prime the cartridge and the pressure measurement tubing, attach a sterile
fluid-filled syringe to the side port of the stopcock.
Laborie NXT Pro Owner's Manual LBL-00146 [C], MAN2010
on page 107.
T-DOC Air-Charged Catheters
Quick Start
on
28