Set The Investigation Protocol; Connect The Catheter - LABORIE Solar Blue Service & Installation Manual

Table of Contents

Advertisement

8.3.3

Set the investigation protocol

In the protocol settings of the stationary measurement program you will find
the channel configuration by clicking on the Connections button. The channel
names will pre-fixed by the Luna module.
The default settings are:
Pves (1)
Pabd (2)
Pura (3)
EMG1
Conductance1
8.3.4

Connect the catheter

Solid state or air charged transducers with Lemo 4 plug can be connected to
the Lemo 4 inputs for measuring pressures. In case you have a catheter with
conductance rings, connect the Lemo 3 plug to the conductance input on the
Luna. The color of the air charged sensors represent the pressures as follows:
Color air charged catheter
Yellow
Blue
Green
The pressure sensors of the air charged catheter must be left in the 'open'
position.
Luna Wireless Module
= Luna P1
= Luna P2
= Luna P3
= Luna EMG
= Luna Conductance
Pressure
Pves
Pabd
Pura
© 2008 – 2020 by MMS B.V.