8.3.3
Set the investigation protocol
In the protocol settings of the stationary measurement program you will find
the channel configuration by clicking on the Connections button. The channel
names will pre-fixed by the Luna module.
The default settings are:
Pves (1)
Pabd (2)
Pura (3)
EMG1
Conductance1
8.3.4
Connect the catheter
Solid state or air charged transducers with Lemo 4 plug can be connected to
the Lemo 4 inputs for measuring pressures. In case you have a catheter with
conductance rings, connect the Lemo 3 plug to the conductance input on the
Luna. The color of the air charged sensors represent the pressures as follows:
Color air charged catheter
Yellow
Blue
Green
The pressure sensors of the air charged catheter must be left in the 'open'
position.
Luna Wireless Module
= Luna P1
= Luna P2
= Luna P3
= Luna EMG
= Luna Conductance
Pressure
Pves
Pabd
Pura