 Remove air from tube.
 Connect the flowmeter via Bluetooth (when the hardware test Bluetooth is
used) or,
 Connect the flowmeter via a wired connection to an MMS-bus receptacle
(when the hardware test Solar (USB) is used). Make sure the power
adapter is connected to the Solar Blue module in this case.
 Place the catheter in the water container which is positioned on the
flowsensor.
 Click the Calibrate pump button. You will see the progress bar "Busy
calibrating". The pump stops automatically after 5 minutes.
 When there is no weight increase during calibration, you will see an error
message ('No weight increase detected on flowmeter'). Click the OK button
to abort calibration.
4.4.3

Pressure test

Choose the TAB Pressures / EMG to test the functioning of the pressure
connectors on the Solar Blue module and the connected transducers or
catheters.
Figure 4.21 Test the pressures