RAPID RESPONSE WATER SYSTEM FOR STERIS OPERATION & MAINTENANCE MANUAL Manufactured With Pride In The USA www.ameriwater.com • 800-535-5585 AmeriWater • 3345 Stop 8 Road • Dayton, OH 45414 98-2024 Rev. A...
TABLE OF CONTENTS THEORY OF OPERATION ..................1 SPECIFICATIONS ..................... 2 SYSTEM FLOW DIAGRAM ..................4 SYSTEM INSTALLATION ..................5 SRO ........................... 5 WALL MOUNT DI ....................... 6 STORAGE TANK ....................... 7 ENDOTOXIN FILTER ....................8 SYSTEM START-UP ....................9 OPERATION ......................
THEORY OF OPERATION This system is designed to pretreat and purify water for use in health care applications. Below is the sequence of the water progression as it goes thru the components of the MOBILE SPD SYSTEM. Each component’s theory of operation is listed in order of water progression thru the system.
REPONSE WATER SYSTEM. Your system was thoroughly tested and in excellent condition when it was shipped to you. However, because damage during shipment is possible, please unpack and carefully inspect the system as soon as you receive it. Please notify AmeriWater if any problems are encountered.
Page 6
Note: For specifications on system components related to electrical requirements, system pressures and voltage draws, please refer to listed component manuals as listed below. For 0021003 Loop Endotoxin Filter, refer to Manual 98-0033 For 00SRO1095 RO, refer to Manual 98-0150 For 00MW200 Silex Wall Mount Deionizer, refer to Manual 98-0160 For 00HC-0090-PRV Healthcare Storage Tank, refer to Manual 098-0001 98-2024 Rev.
SYSTEM INSTALLATION CAUTION: Local plumbing and electrical codes must be observed. The electrical source for the S100 controller must be a single phase, 3-conductor receptacle with a ground fault interrupter (GFI) at 120V, 20amp, and 60Hz. The proper polarity and ground integrity must be initially checked and thereafter maintained.
WALL MOUNT DI The Ameriwater Wall Mount Deionizer should be installed on a wall capable of supporting 50 pounds. 1. Locate the DI Polisher to the right of the SRO, just before the storage tank and within 6 feet from a 120 volt 60 hertz receptacle 2.
5. Make sure that the deionizing cartridges are installed in the filter housing and that the bowls are hand tightened only. 6. Plug transformer into the receptacle and open the inlet and sample port valves. STORAGE TANK Locate the storage tank on a firm, level foundation and anchored securely. Connect the Wall Mount DI product outlet hose to the storage tank’s inlet.
Run the float switch quick disconnect from storage tank to the SRO controller and lock in place. (Figure 2.4) ENDOTOXIN FILTER Locate and install the Endotoxin-Retentive Filter assembly on or near the wall. Connect the 3/4” pump discharge hose to inlet valve on the endotoxin filter assembly. (Figure 2.3) 98-2024 Rev.
3.0 SYSTEM START-UP WARNING: This Reverse Osmosis System (RO) contains a preservative solution to prevent microbiological growth and freezing. Discard all product water for at least two hours of operation before placing the RO in service. NOTE: As each component is started and flushed of any debris, make sure to divert loop water to the drain to flush loop of any debris as well.
BYPASS PROCEDURE This system has a bypass option. The bypass can be used to repair or maintenance any part of the system. Use the option of bypassing the system, when you have to remove or maintenance any component of your system (RO, Silex or Storage Tank Distribution Pump).
MONITORING Under AAMI TIR34 requirements, the below listed items need to be monitored on varying intervals. 1. Ultra filters are to be monitored daily for a pressure drop. The pressure drop across the filter that indicates the need for replacement is 15 PSI. 2.
Page 15
Fill out the monitoring log daily. Having this information available will help to quickly diagnose issues related to performance. Failure to carry out the daily monitoring and maintenance at the indicated intervals will result in reduced performance of the SRO system and may void the warranty.
PREVENTATIVE MAINTENANCE SCHEDULE FREQUENCY For 00SRO1095 RO, Pre-filter, part number 20-5101, must be replaced in the event of listed instances below: 1. High chlorine levels of greater than 0.1ppm 2. Pressure drop across the filter of greater than 15PSI. Membrane, part number R22-2521, must be replaced in the event of listed instances below: 1.
WINTERIZATION Step 1 – Disconnect all Power sources. Filter housings- Open the sample ports to all filter housings in the system (i.e. RO prefilter, Wall Mount DI, Endotoxin filters) to relieve water pressure. Open filter housings, remove filters cartridges and discard cartridges. Undo any plumbing connections to the filter housing that will not allow water to freely drain from the fitting.
( Prefilter, Vent Filter and DI ) 21530233 O-RING,FLTR HSG,10IN BB,SQ THD ( Loop Endotoxin Filter Assembly) *Call AmeriWater or your AmeriWater distributor for pricing. For 00SRO1095 RO, refer to Manual 98-0150 For 00MW200 Silex Wall Mount Deionizer, refer to Manual 98-0160...