Monitoring the oil cleanliness when filling machines and systems ................. 23 10.6 Monitoring the oil cleanliness when cleaning machines and systems in the bypass flow ..........24 Page 2 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
Page 3
Technical data ................................36 17.3 Operating conditions ..............................37 17.4 Hydraulic circuit diagram ............................... 37 Appendix ..................................38 18.1 Declaration of conformity ............................. 38 18.2 Spare parts ..................................39 Page 3 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
Indicates a dangerous situation which may result in light to moderate injury if not avoided. NOTE Indicates property damage: The product or surrounding could be damaged. Table 2: Hazard classes according to ANSI Z536.6-2006 Page 4 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
In this documentation the following terms are used: Term Meaning Table 4: Terms 1.3.4 Abbreviations In this documentation the following abbreviations are used: Term Meaning UMPCL Pumping unit with condition monitoring Table 5: Abbreviations Page 5 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
Therefore only use this product in safety-related applications if this use is explicitly specified and permitted in the product documentation, e.g. in explosion protection areas or in safety-related parts of a control system (functional safety). ARGO-HYTOS assumes no liability for damages resulting from improper use. The risks associated with improper use are solely with the user.
› Observe all instructions on the product. › People who assemble, operate, disassemble or maintain ARGO-HYTOS products may not do so under the influence of alcohol, other drugs or medications that affect the responsiveness. › Only use manufacturer-approved accessories and spare parts, in order to prevent personal danger due to unsuitable spare parts.
Do not use aggressive cleaning agents for cleaning. Clean the product with a suitable cleaning fluid. › Do not use a high pressure cleaner. › Do not use compressed air to clean function interfaces such as seal areas. Page 8 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
On the front of the unit there is a cable holder, on which the 230V or 400V power cable can be wound up for better transport in case of non-use. Page 10 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
Page 11
A missing protective strainer may lead to destruction of the pump. Checking / changing the suction filter is described in "11.4 Checking / changing the suction filter element (pump protection filter)" on page 27. Page 11 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
The 3-way ball valve makes it possible to determine whether the oil cleanliness monitoring is to be performed before or after the filter. Potentiometer (only available in the adjustable flow version) The potentiometer is used for manual adjustment of the flow rate in the range of 20-70 l/min. Page 13 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
Identification of the product Producer address Manufacturing Type date (encrypted) Filter element type Nominal volume flow Fineness of the filter Maximum operating pressure Serial number Fig.6: Nameplate Page 14 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
80 %. Before storage over a period of more than 6 months, the device should be filled with oil in order to preserve it against corrosion. Page 15 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
The following steps are to be followed in detail: › connect the 230V (or 400V) voltage plug to the local power supply. Page 16 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
The ventilating valve (1) can be checked by means of manual venting (red cap / 2). After repeated actuation of the valve, oil should escape. This check also prevents the float valve from being glued. Fig.8: Venting Page 17 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
COM port, it is possible to use a USB-serial converter. NOTE The software can be downloaded from www.argo-hytos.com. The stored results are transferred to the connected devices such as PC, laptop, tablet or smartphone and can be stored there. Page 18 www.argo-hytos.com...
Thus, the residual oil from the filter housing and the lines above the oil level is fed back into the same container via the discharge pipe. 9. Fix the suction lance / suction pipe and the discharge pipe to the corresponding brackets at the filter unit. Page 20 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
Make sure that the device is properly vented. › An exact determination of the cleanliness class is possible in a viscosity range of 15 mm /s to 300 mm Page 21 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
1. During the pumping or filtration process, the output flow of the unit can be regulated in the range of 20-70 l/min. 2. In order to increase the flow rate, turn the potentiometer knob clockwise. Fig. 12: Front panel with potentiometer for flow adjustment Page 22 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
/ s to 300 mm / s. › When using the adjustable flow version, it is recommended to measure the cleanliness class at a nominal flow rate of 45 l/min Page 23 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
10. Switch off the filter unit at the circuit breaker. 11. Fix the suction lance / suction pipe and the discharge pipe to the corresponding brackets at the filter unit. Page 24 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
Contact temperatures according to DIN EN563 (3) and DIN EN13202 (4) may be exceeded during operation. › Allow the filter unit to cool down before touching it. Page 25 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
2. Attach the filter element in arrow direction 1 and lock it in arrow direction 2. Fig. 19: Attaching the filter element Page 26 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
Fig. 20: Installing the filter element NOTE ARGO-HYTOS offers you a comprehensive service package for the repair of the filter unit UMPCL 045. › Repairs to the filter unit UMPCL 045 may only be carried out by the manufacturer or his authorized dealers and branches.
3. Clean filters using cleaning agent and compressed air. If necessary install new filters (order no. 15077600) 4. Install filter to the system Fig. 22: Removing screen filters Page 28 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
The filter unit UMPCL 045 is a component which does not have to be taken out of service. Therefore, the chapter in this manual does not contain any information. Page 29 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
13. Disassembly This chapter does not contain any information for your device. Page 30 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
Therefore, dispose of the filter unit and the pressure fluid according to the national regulations of your country. Dispose of pressure fluid residues according to the relevant safety data sheets for these pressure fluids. Page 31 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
Do not modify the UMPCL 045 filter unit. NOTE The warranty of ARGO-HYTOS applies to the delivered configuration and extensions, which were taken into account during configuration. The warranty expires after a conversion or extension that goes beyond the conversions or extensions described here.
Get a clear impression about the cause of trouble. Possibly consult the direct operator. If you cannot correct the error, please contact one of the contact addresses listed under www.argo-hytos.com. If the device should not accept any more entries, turn it off briefly and then turn it on again after a few seconds. After initialization, the particle monitor will operate as usual.
Page 34
Upper limit value: 24 (for 4, 6 and 14 μm) improvement (e.g. Lower limit value: 10 (for 4, 6 and 14 μm) order number 10 at 4, 6 and 14 μm). Table 8: Error description Page 34 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
73 (Under operating conditions permissible for continuous operation) max. 78 (Under operating conditions permissible for short-term operation) Dimensions l x w x h 786 x 564 x 1,045 Table 9: Technical data Page 36 www.argo-hytos.com Subject to change · UMPCL 045_41733800 · 0321 · EN...
18. Appendix 18.1 Declaration of conformity EC Declaration of Conformity We, the company, ARGO-HYTOS Polska sp. z o.o. Władysława Grabskiego 27 32-640 Zator, Poland, declare on our sole responsibility that the products in the model series Mobile filter unit UMPCL 045...