ELECTING AN OUTPUT PORT AND WAVELENGTH
52

Selecting an output port and wavelength

1.
To select output, tap SM or MM on top of the display.
– or –
[A]
Press the
key, tap [Output
output.
The menu is closed and the measurement display is shown.
2.
To select a wavelength, tap the desired wavelength(s) on the
measurement display.
– or –
[A]
Press the
key, tap [Source
wavelength(s).
If Auto-
is not selected, only one wavelength can be selected.
λ
To select Auto-λ see
[F]
3.
Press the
key.
4.
To turn on the selected lasers, tap the [Laser
or press the central key.
About Auto-λ mode
When in one of the Auto-λ modes, the ORL-85/ORL-85P
embeds wavelength ID information into the emitted signal(s).
This wavelength ID information can be utilized by a compatible
Viavi power meter (e.g. OLP-85) for automatic wavelength setup
and for simultaneous reception of multiple wavelengths.
The ORL-85/ORL-85P provides two Auto-λ modes:
• Multi-λ: Multi-λ is a proprietary Viavi solution that reduces
testing time to a minimum by simultaneously testing at both
wavelengths. When activated both wavelength are measured
and displayed simultaneously.
Auto-λ-Multi should be selected every time the optical
powermeter supports the Auto-λ-Multi feature.
• Serial-λ automatically toggles between the selected
wavelengths. Combined with a compatible power meter (e.g.
OLP-3x, OLP-85), the power meter automatically sets up with
correct wavelength settings.
Serial-λ should be selected when low levels are expected on
the optical power meter or when the power meter doesn't
support Multi-λ.
Selection]
Selection]
"About Auto-λ μοδε
and tap the desired
and tap the desired
.
On/Off]
button