Watson-Marlow 730 U Installation, Operating And Maintenance Manual

Table of Contents

Advertisement

Advertisement

Table of Contents
loading

Summary of Contents for Watson-Marlow 730 U

  • Page 1: Table Of Contents

    730 U/UN Installation, operating, and maintenance manual Contents 1 Declaration of conformity 2 Declaration of incorporation 3 When you unpack your pump 3.1 Unpacking your pump 3.2 Packaging disposal 3.3 Inspection 3.4 Components supplied 3.5 Storage 4 Information for returning pumps...
  • Page 2 11.2 Wiring the US NEMA module 11.3 Earth screening of control cables on NEMA module 12 Start-up check list 13 Control wiring 13.1 Upper D-connector 13.2 Lower D-connector 13.3 Standard - 25 way D: Inputs and Outputs 13.4 Standard - 25 way D: power supply limits 13.5 N - Modules Standard and SCADA 13.6 Standard N - Module 13.7 SCADA N-Module...
  • Page 3 20.3 Resume interrupted doses 20.4 Master dose 20.5 Manual dosing 21 Main menu 21.1 Security settings 21.2 General settings 21.3 Control settings 21.4 Configure outputs 21.5 Configure inputs 21.6 Help 22 Troubleshooting 22.1 Leak detection 22.2 Error codes 22.3 Technical support 23 Drive maintenance 24 Drive spares 25 Pumphead replacement...
  • Page 4 Original instructions The original instructions for this manual have been written in English. Other language versions of this manual are a translation of the original instructions m-730un-en-08.1...
  • Page 5: Declaration Of Conformity

    7. Intertek Testing and Certification Ltd, No: 3272281, performed compliance testing to BS EN 61010-1:2010, IEC 61010-1:2010, UL 61010-1:2010 and CAN/CSA C22.2 Bo 61010-1:2010 and issued certification of compliance to these standards. Signed for and behalf of: Watson Marlow Ltd Falmouth, November 2019 Simon Nicholson, Managing Director, Watson-Marlow Limited m-730un-en-08.1...
  • Page 6: Declaration Of Incorporation

    The pump head is incomplete and must not be put into service until the machinery into which it is to be incorporated has been declared in conformity with the provisions of the Directive. Person authorised to compile the technical documents: Nancy Ashburn, Head of Design & Engineering, Watson-Marlow Ltd Place and date of declaration: Watson-Marlow Ltd, 20.04.2020 Responsible person: Simon Nicholson, Managing Director, Watson-Marlow Ltd m-730un-en-08.1...
  • Page 7: When You Unpack Your Pump

    When you unpack your pump Unpacking your pump Unpack all parts carefully, retaining the packaging until you are sure all components are present and in good order. Check against the components supplied list, below. Packaging disposal Dispose of packaging materials safely, and in accordance with regulations in your area. The outer carton is made of corrugated cardboard and can be recycled.
  • Page 8: Information For Returning Pumps

    Information for returning pumps Before returning products, they must be thoroughly cleaned/decontaminated. The declaration confirming this should be completed and returned to us in advance of the item being shipped. You are required to complete and return a decontamination declaration stating all fluids that have been in contact with the equipment being returned to us.
  • Page 9: Warranty

    Watson-Marlow has been advised of the possibility of such damages. Watson-Marlow shall not be liable for any loss, damage, or expense directly or indirectly related to or arising out of the use of its products, including damage or injury caused to other products, machinery, buildings, or property. Watson- Marlow shall not be liable for consequential damages, including without limitation, lost profits, loss of time, inconvenience, loss of product pumped, and loss of production.
  • Page 10: Safety Notes

    If the pump is used in a manner not specified by Watson-Marlow Ltd, the protection provided by the pump may be impaired. Any person who is involved in the installation or maintenance of this equipment should be fully competent to carry out the work.
  • Page 11 Do not fit any devices to the drive unit other than those tested and approved by Watson-Marlow. Doing so could lead to injury to persons or damage to property for which no liability can be accepted.
  • Page 12 Please refer to the chemical compatibility guide which can be found at: www.wmftg.com/chemical. If you need to use the pump with any other chemical please contact Watson-Marlow to confirm compatibility. If the Auto restart feature is enabled it may cause the pump to start as soon as power is turned on.
  • Page 13: Pump Specifications

    Pump specifications Specification ratings Operating temperature 5C to 40C (41F to 104F) Storage temperature 730: -25C to 65C (-13F to 149F) Humidity (non-condensing) 80% up to 31C (88F) decreasing linearly to 50% at 40C (104F) Maximum altitude 2000m (6560ft) Power Rating 730: 350VA Supply voltage 100-120V/200-240V 50/60Hz 1pH (Subject to regional cord sets and...
  • Page 14: Pumphead Options

    Pumphead options 730 pump range 720R, 720R/RX, 720RE, 720RE/REX: m-730un-en-08.1...
  • Page 15: Good Pump Installation Practice

    Good pump installation practice General recommendations It is recommended that the pump is sited on a flat, horizontal, rigid surface, free from excessive vibration, to ensure the correct lubrication of the gearbox and correct pumphead operation. Allow a free flow of air around the pump to ensure that heat can be dissipated. Ensure that the ambient temperature around the pump does not exceed the recommended maximum operating temperature.
  • Page 16: Dos And Don'ts

    If in any doubt about the compatibility of a tube material and the duty fluid, request a Watson-Marlow tube sample card for immersion trials. When using Marprene or Bioprene continuous tubing, do re-tension the tube after the first 30 minutes of running.
  • Page 17: Pump Operation

    Pump operation 10.1 Keypad Layout and Key IDs Change value / Colour TFT display move selection bar Mode Home Direction of flow Function keys Stop Start HOME key When the HOME key is pressed it will return the user to the last known operating mode. If modifying pump settings when the HOME key is pressed, it will disregard any setting changes and return you to the last known operating mode.
  • Page 18: Starting And Stopping

    10.2 Starting and stopping 10.3 Using up and down keys 10.4 Maximum speed 10.5 Change rotation direction m-730un-en-08.1...
  • Page 19: Connecting To A Power Supply

    Connecting to a power supply A well regulated electrical mains supply is required along with cable connections conforming to the best practice of noise immunity. It is not recommended to site these drives alongside electrical devices that may generate mains-borne noise, for example 3-phase contactors and inductive heaters. Set the voltage selector to 115V for 100-120V 50/60Hz supplies or 230V for 200- 240V 50/60Hz supplies.
  • Page 20: Conductor Colour Coding

    11.1 Conductor colour coding Conductor type European colour North American colour Line Brown Black Neutral Blue White Ground Green/Yellow Green 11.2 Wiring the US NEMA module The NEMA 4X modules fitted to 530, 630 and 730 Du, S, U and PROFIBUS cased pumps with part numbers ending with an A (xxx.xxxx.xxA) have two pairs of wiring ports.
  • Page 21: Start-Up Check List

    Start-up check list Note: See also "Tube replacement " on page 113. Ensure that proper connections are achieved between the pump and suction and discharge piping. Ensure proper connection has been made to a suitable power supply. Ensure that the recommendations in the section "Good pump installation practice " on page 15 are followed.
  • Page 22: Control Wiring

    Control wiring Never apply mains power to the D-connectors. Apply the correct signals to the pins shown. Limit signals to the maximum values shown. Do not apply voltage across other pins. Permanent damage, not covered by warranty, may result. Keep 4-20mA and low voltage signals separate from mains power. Use separate glanded input cables.
  • Page 23 Note: Not all functions may be available, those available will depend on your pump model. Key to symbols DRY (NO LEAK) INPUT WET (LEAK DETECTED) OUTPUT STOP CLOCKWISE ROTATION KEYPAD DIRECTION CHANGE MANUAL (KEYPAD) CONTROL REMOTE DIRECTION CHANGE ANALOGUE (4-20mA/0-10V) CONTROL ANTI-CLOCKWISE ROTATION ALARM PULL UP START DOSE ON RISING EDGE...
  • Page 24 Input or Signal Name Configurable Signal response output AUTO/MAN AUTOMAN AUTOMAN LOGIC OUT 1 LOGIC OUT 1 Output 1 LOGIC OUT 2 LOGIC OUT 2 Output 2 m-730un-en-08.1...
  • Page 25 Input or Signal Name Configurable Signal response output LOGIC OUT 3 LOGIC OUT 3 Output 3 LOGIC OUT 4 logic_out4.pdf LOGIC OUT 4 Output 4 COMMON Commoning terminals, used for voltage supply for logic outputs m-730un-en-08.1...
  • Page 26 Wiring the lower D connector Input Function Configurable Signal response output TACHO FREQUENCY TACHO F Hz TACHO 4-20mA TACHO mA TACHO 0-10V 0-10V m-730un-en-08.1...
  • Page 27 Input Function Configurable Signal response output TTL 1 Output 1 TTL OUT 1 TTL 2 Output 2 TTL OUT 2 DIRECTION DIRECTION DIRECTION m-730un-en-08.1...
  • Page 28 Input Function Configurable Signal response output START/STOP START/STOP Stop = high Stop = low START/STOP DIRECTION ENABLE DIRECTION ENABLE DIRECTION ENABLE m-730un-en-08.1...
  • Page 29 Input Function Configurable Signal response output DOSE DOSE DOSE ANALOG 1 Either 0-10V or 4-20mA ANALOG 1 0-10V 10V REF ANALOGUE 1 m-730un-en-08.1...
  • Page 30: Standard - 25 Way D: Power Supply Limits

    Input Function Configurable Signal response output ANALOG 2 ANALOG 2 13.4 Standard - 25 way D: power supply limits Signal Connector Pin numbers Voltage Load 5V Ref Upper 18, 19 5V @ no load Total of all pins 10mA max Lower 9, 19, 23 12V Ref...
  • Page 31: Standard N - Module

    13.6 Standard N - Module Note: Not all functions may be available, those available will depend on your pump model. Tacho o/p RS485 Relay 4 Relay 3 Relay 2 Relay 1 Spare supplies Commoning Freq -12V +12V (maximum +24VDC) (Planned) (planned) +12V +10V...
  • Page 32 Key to symbols DRY (NO LEAK) INPUT WET (LEAK DETECTED) OUTPUT STOP MANUAL (KEYPAD) CONTROL CLOCKWISE ROTATION KEYPAD DIRECTION CHANGE ANTI-CLOCKWISE ROTATION ANALOGUE (4-20mA/0-10V) CONTROL REMOTE DIRECTION CHANGE EXTERNAL VOLTAGE INPUT START DOSE ON RISING EDGE Standard N Module: Input/output Connectors Connector No.
  • Page 33 Connector No. Function Input or Configurable Signal response output DOSE DOSE STOP HIGH Stop = high RUN /STOP STOP LOW RUN /STOP Stop = low RUN /STOP ANALOGUE 1 DIRECTION EN m-730un-en-08.1...
  • Page 34 Connector No. Function Input or Configurable Signal response output 10V 10mA Max REMOTE POT ANALOGUE 2 Leak = high LEAK DETECT HIGH LEAK DETECT LEAK DETECT LOW LEAK DETECT Leak = low LEAK DETECT R S 4 8 5 0 V A B J 1 0 Hz=5V TTL V=0-10V...
  • Page 35 Connector No. Function Input or Configurable Signal response output Sum of all 12V currents = 10mA SPARE SUPPLIES DC Max COMMONING J14,15,16,17 LOW=N/C HIGH=N/O RELAY 1-4 24V DC MAX Switches in or out a 124R termination resistor - May be required for long cable lengths Switches in or out a filter capacitor for optional use when...
  • Page 36: Scada N-Module

    13.7 SCADA N-Module Note: Not all functions may be available, those available will depend on your pump model. RS485 Tacho o/p Relay 4 Relay 3 Relay 2 Relay 1 +12V +10V Direction Auto/man i/p Tube monitor Analog 2 Rem pot Direction i/p enable link Analog 1 Run/stop i/p Dose i/p...
  • Page 37 Key to symbols INPUT DRY (NO LEAK) WET (LEAK DETECTED) OUTPUT STOP CLOCKWISE ROTATION KEYPAD DIRECTION CHANGE MANUAL (KEYPAD) CONTROL ANALOGUE (4-20mA/0-10V) CONTROL REMOTE DIRECTION CHANGE ANTI-CLOCKWISE ROTATION AC (ALTERNATING CURRENT) INPUT START DOSE ON RISING EDGE SCADA N Module: Input/output Connectors Connector No.
  • Page 38 Connector No. Function Input or Configurable Signal Response Output ANALOGUE 1 DIRECTION EN 10V 10mA Max REMOTE POT ANALOGUE 2 Leak = high LEAK DETECT LEAK DETECT Leak = low 0-10V 4-20mA TACHO O/P m-730un-en-08.1...
  • Page 39 This is indicated below and the relevant bytes are listed in the table. No further changes should be made to the GSD file and Watson-Marlow accepts no responsibility for the pump failures arising from the changes to the GSD file.
  • Page 40 8 bit Byte 3 Min speed (high byte of 16-bit unsigned) 8 bit Byte 4 Min speed (low byte of 16-bit unsigned) 8 bit Byte 5 Max speed (high byte of 16-bit unsigned) 8 bit Byte 6 Max speed (low byte of 16-bit unsigned) 8 bit Byte 7 Fail safe mode configuration...
  • Page 41 Pumphead type Value Pumphead 0x—80 720R 0x—81 720RE (Sanitary element) 0x—82 720RE (Industrial element) Note: Numerical values cannot include a decimal point. To enter a speed, move the decimal point one place to the right and enter an integer. To enter a tube size, move the decimal point two places to the right and enter an integer.
  • Page 42 PROFIBUS Ident: 0x0F70 GSD file: WAMA0F70.GSD Config: 0x62, 0x5D (3 words out, 14 words in) User parameter bytes: Cyclic data write (from master to pump) 16 bit Byte 1 Control word (high byte) Byte 2 Control word (low byte) 16 bit Byte 3 Pumphead speed setpoint (unsigned) (high byte) Byte 4...
  • Page 43 Cyclic data read (from pump to master) 32 bit Byte 7 Tacho counter (high byte) Byte 8 Tacho counter (low byte) Byte 9 Tacho counter (high byte) Byte 10 Tacho counter (low byte) 16 bit Byte 11 Pump outputs flow calibration in deci μl (For example: A value of 4 indicates 40μl) (high byte) Byte 12 Pump outputs flow calibration in deci μl (For example: A value of 4...
  • Page 44 Status word Guard (Guard enabled models only) (1=guard open) Over current error Under voltage error Over voltage error Over temperature error Motor stalled Tacho fault Leak detected Low setpoint - out of range High setpoint - out of range Reserved Reserved Reserved Device Related Diagnostic Data...
  • Page 45 Channel related diagnostic data Byte 1 Header Byte 2 Channel type Byte 3 Channel related error code Channel related diagnostic data Byte 3 Global error =0xA9 (General error) Over current =0xA1 (Short circuit) Under voltage =0xA2 (Under voltage) Over voltage =0xA3 (Over voltage) Over temp =0xA5 (Over temp)
  • Page 46 Avoid sharp bends in PROFIBUS communication cables. Please follow PROFIBUS wiring standards. m-730un-en-08.1...
  • Page 47: Switching The Pump On For The First Time

    Switching the pump on for the first time Power up the pump. The pump displays the start-up screen with the Watson-Marlow Pumps logo for three seconds. 14.1 Selecting the display language 1. Use the ∧/∨ keys to select your chosen language, and press SELECT.
  • Page 48 2. Your selected language will now be displayed on screen. Choose CONFIRM to continue. All text will now appear in your chosen language. 3. Choose REJECT to return to the language choice screen. This then proceeds to the home screen. m-730un-en-08.1...
  • Page 49: First-Time Start-Up Defaults

    14.2 First-time start-up defaults The pump is preset with operational parameters as shown in table below. Parameter 730 default Language Not set Default mode Manual Default manual speed 360rpm Pump status Stopped Max speed 360rpm Direction Pumphead 720R Tube size 25.4mm Tube material Bioprene...
  • Page 50 Parameter 730 default Output 2 - status High = CW Output 3 Auto/man Output 3 - status High = auto Output 4 General alarm Output 4 - status High = alarm The pump is now ready to operate according to the defaults listed above. Note: The display background colour changes according to running state as follows: White background indicates pump stopped Grey background indicates pump running...
  • Page 51: Switching The Pump On In Subsequent Power Cycles

    Switching the pump on in subsequent power cycles Subsequent power-up sequences will jump from the start-up screen to the home screen. The pump runs a power-on test to confirm proper functioning of the memory and hardware. If a fault is found, an error code is displayed. The pump displays the start- up screen with the Watson- Marlow Pumps logo for three seconds followed by the home screen Start-up defaults are those in place when the pump was switched off last...
  • Page 52: Mode Menu

    Mode menu Press MODE to display the Change mode menu. Use the ∧ and ∨ keys to scroll through the available modes. • Manual(default) • Flow calibration • Analog • Memodose • CANCEL Use SELECT to choose mode. Use the right hand function key to alter mode settings. m-730un-en-08.1...
  • Page 53: Manual

    Manual All settings and functions of the pump in manual mode are set and controlled by means of key-presses. Immediately after the start-up display sequence detailed in: "Switching the pump on in subsequent power cycles " on page 51, the manual mode home screen will be displayed unless auto restart is enabled.
  • Page 54: Stop

    17.2 STOP Stops the pump. The display background changes to white. If the pump is not running pressing this has no effect. 17.3 INCREASING AND DECREASING FLOW RATE Using the ∧ and ∨ keys will increase or decrease the flow rate. Decreasing flow rate A single key press will decrease flow rate by the least significant digit of the chosen flow rate unit.
  • Page 55: Max Function (Manual Mode Only)

    17.4 MAX FUNCTION (Manual mode only) Press and hold the MAX key to run at maximum flow. Release the key to stop the pump. The volume dispensed and time elapsed are displayed while the MAX key is pressed and held. m-730un-en-08.1...
  • Page 56: Flow Calibration

    Flow calibration This pump displays flow rate in ml/min. 18.1 Setting the flow calibration Using the ∧ /∨ keys, scroll to Flow calibration and press CALIBRATE. Using the ∧ /∨ keys, enter the maximum flow rate limit and press ENTER. m-730un-en-08.1...
  • Page 57 Press START to begin pumping a volume of fluid for calibration. Press STOP to stop pumping fluid for the calibration. Use the ∧ /∨ keys to enter the actual volume of fluid pumped. m-730un-en-08.1...
  • Page 58 To accept the new calibration press ACCEPT or RE-CALIBRATE to repeat the procedure. Press HOME or MODE to abort. The pump is now calibrated. m-730un-en-08.1...
  • Page 59: Analog Mode

    Analog mode In this remote operating mode the flow rate will be proportional to the external milliamp or voltage signal input received by the pump. The relationship between the external analog signal and the flow rate is determined by configuring the two points A and B as shown in the graph below. The rate of flow can be proportional or inversely proportional to the analog input.
  • Page 60: Analog Calibration

    19.1 Analog calibration The pump must be stopped before attempting to calibrate the values. High and low signals must be within range. If the signal sent is out of range you will not be able to set the signal input value and progress to the next step in the process. Select MENU, then select MODE menu.
  • Page 61: Calibrate Input

    19.2 Calibrate Input 1 Using the ∧ /∨ keys, scroll to Analog Input and press SELECT. Select the input signal type using the ∧ /∨ keys and press SELECT. The pump gives the option to enter the high and low mA or V signals manually or via the analog input. Entering mA values is described here, however the process for entering V signals is identical.
  • Page 62: Setting A High Signal

    Choose whether to enter the current values manually via the keypad, or whether to apply the current signals electrically to the analog input. 19.3 Setting a high signal Send the high signal input to the pump, or enter the current value using the ∧ / ∨ keys. m-730un-en-08.1...
  • Page 63: Setting High Flow Calibration

    ACCEPT appears when high mA signal is within tolerance limits. Press ACCEPT to accept the high signal input or CANCEL to return to the previous screen. 19.4 Setting high flow calibration Using the ∧ / ∨ keys, scroll to the desired flow rate. Select SET FLOW or press BACK to return to the previous screen.
  • Page 64: Setting A Low Signal

    19.5 Setting a low signal Send the low signal input to the pump, or enter the current value using the ∧ / ∨ keys. If the range between the low and high signal is less than 1.5mA, the following error message will be displayed.
  • Page 65: Setting Low Flow Calibration

    ACCEPT appears when low mA signal is within tolerance limits. Press ACCEPT to accept the low signal input or CANCEL to return to the previous screen. 19.6 Setting low flow calibration Using the ∧ / ∨ keys, scroll to the desired scale factor. Select SET FLOW. m-730un-en-08.1...
  • Page 66 This proceeds to the screen confirming calibration is complete. Select ANALOG to start in analog mode or MANUAL to continue in manual mode. m-730un-en-08.1...
  • Page 67: Memodose Mode

    MemoDose mode Each time the pump is started by pressing START, it records the number of pumphead revolutions which occur until STOP is pressed. The number of revolutions is proportional to the volume of fluid which has been dispensed: the dose. MemoDose mode allows the user to repeat-dose a precise volume of fluid. This can be achieved by dispensing a master dose, or entering a dose volume manually using the keypad.
  • Page 68: Resume Interrupted Doses

    20.3 Resume interrupted doses MemoDose Mode gives you the opportunity to resume interrupted doses following a power cycle (Note: Auto-restart must be on for this to function). Alternatively interrupted doses can be discarded and a new dose started when power is resumed. From the MemoDose settings screen, use the ∧...
  • Page 69: Master Dose

    20.4 Master dose Using the ∧ /∨ keys, scroll to Master dose and press SELECT. The following screen is displayed; press MANUAL to enter a dose via the keypad or DOSE to dispense a master dose. m-730un-en-08.1...
  • Page 70 Dispensing a master dose Press START to begin dispensing the Master Dose. Press STOP to finish dispensing the Master Dose. m-730un-en-08.1...
  • Page 71 Manual dose entry Press MANUAL on the master dose screen. Using the ∧ /∨ keys enter the required dose volume and press SELECT to record the master dose or CANCEL to return to MemoDose Settings. Save dose volume Press SAVE to record the Master Dose or CANCEL to return to MemoDose settings. m-730un-en-08.1...
  • Page 72: Manual Dosing

    When MemoDose SETTINGS are complete the following screen is displayed; press MEMODOSE to start in MemoDose Mode or BACK to return to MemoDose settings. 20.5 Manual dosing From the MemoDose Home screen, press START to deliver a dose. The screen displays the dose flow rate and the dose remaining counting down from 100% to 0%.
  • Page 73 If the delivered dose differs from the volume required, the percentage may be adjusted within the limits 50% to 150% of the master dose. Use the ∧ /∨ keys to alter the percentage. The new dose size is displayed as a percentage on the home screen. 107% If STOP is pressed during dosing, the pump stops.
  • Page 74: Main Menu

    Main menu To access the main menu press the MENU button from one of the HOME screens or INFO screens. This will display the main menu as shown below. Use the ∧ /∨ keys to move the selection bar between the available options.
  • Page 75 m-730un-en-08.1...
  • Page 76: Security Settings

    21.1 Security settings Security settings can be changed by selecting SECURITY SETTINGS from the Main menu. Auto keypad lock Press ENABLE/DISABLE to switch on/off the Auto keypad lock. When active the keypad will 'lock' after 20 seconds of inactivity. Once locked it will display the screen below when any key is pressed. To unlock the keypad press the two UNLOCK keys together.
  • Page 77 The padlock icon will appear on the operating mode home screen to show that keypad lock is activated. Note that the STOP key will always work whether the keypad is locked or not. PIN protection Using the ∧ / ∨ keys select PIN protection from the SECURITY SETTINGS menu and press ENABLE/DISABLE to switch on/off the PIN protection.
  • Page 78 Now press CONFIRM to check that the number entered is the PIN you require. Press CHANGE to return to PIN entry. m-730un-en-08.1...
  • Page 79 The following screen will be displayed to indicate that the Master PIN has been applied to access all functionality. Press NEXT to selectively enable functionality access for User 1 and User 2. Configure User 1 security settings The PIN PROTECTION level screen will be displayed with User 1 highlighted, press ENABLE to configure User 1 security settings or scroll to configure an alternative User.
  • Page 80 ENABLE user 1 security settings displays the PIN entry screen for User 1. To define a four digit User 1 PIN, use the ∧ /∨ keys to select each digit from 0-9. Once you have the required digit press the NEXT DIGIT key.
  • Page 81 To define the allowed functionality, use the ∧ /∨ keys to select the functionality and press ENABLE. User 1 PIN will allow access to only the enabled functionality, to disable functionality, highlight the enabled functionality and press DISABLE. When all the required functionality has been enabled, press FINISH. m-730un-en-08.1...
  • Page 82 Configure User 2 security settings The PIN PROTECTION level screen will be displayed with User 2 highlighted, press ENABLE to configure User 2 security settings or scroll to configure an alternative User. ENABLE user 2 security settings displays the PIN entry screen for User 2. To define a four digit User 2 PIN, use the ∧...
  • Page 83 To define the allowed functionality, use the ∧ /∨ keys to select the functionality and press ENABLE. User 2 PIN will allow access to only the enabled functionality, to disable functionality, highlight the enabled functionality and press DISABLE. When all the required functionality has been enabled, press FINISH. m-730un-en-08.1...
  • Page 84 Note: Once Security Settings for User 1 and User 2 have been set by the Master, only the Master PIN will allow access to Security Settings. The HOME screen will be displayed. A PIN is now required to access all functionality. The Master PIN accesses all pump functionality and the User 1 and User 2 PINs access only the defined functionality.
  • Page 85 If an incorrect PIN has been entered the following screen will be displayed. NOTE: this screen will also display if the PIN entered does not allow access to that functionality. If a PIN number is entered that is already in use, the following screen will be displayed, press CHANGE to input an alternative PIN or EXIT to abort m-730un-en-08.1...
  • Page 86 If the PIN entered does not allow access to the functionality the following screen will be displayed. Keypad beep From SECURITY settings scroll to Keypad beep using the ∧ /∨ keys and select ENABLE. The pump will now beep at every key press. m-730un-en-08.1...
  • Page 87: General Settings

    PIN entry on start-up The setting PIN entry during start-up can be used to configure the software to choose if PIN entry is required during start up. This feature also means that auto-restart capability is now independent of PIN code entry after start up. If this setting is enabled ü...
  • Page 88 Do not use auto restart for more than 12 times in 24 hours. We recommend remote control where a high number of starts is required. If the Auto restart feature is enabled it may cause the pump to start as soon as power is turned on.
  • Page 89 Use the ∧ /∨ keys to move the selection bar over the required flow unit, then press SELECT. All flow rates displayed on screens will now be in the chosen units. If a mass flow unit is selected, the specific gravity of the fluid must be entered. The following screen is displayed.
  • Page 90 SELECT. If a pump label has been previously defined, this will be displayed on screen to allow editing, otherwise it will display the default label “WATSON-MARLOW". Use the ∧ /∨ keys to scroll through the available characters for each digit. The available characters are 0-9, A-Z and SPACE.
  • Page 91 Press FINISH to save the entry and return to the general settings menu. Pumphead type Select GENERAL SETTINGS from the main menu. Use the ∧ /∨ keys to move the selection bar over Pumphead type and press SELECT. The following screen will be displayed.
  • Page 92 Use the ∧ /∨ keys to move the selection bar over Pumphead and press SELECT. Use the ∧ /∨ keys to move the selection bar over the required pumphead type and press SELECT. m-730un-en-08.1...
  • Page 93 Tube size and tube material Select Tube size from GENERAL SETTINGS, then use the ∧ /∨ keys to move the selection bar over Bore size and press SELECT. Use the ∧ /∨ keys to move the selection bar over the tube size to be used and press SELECT. m-730un-en-08.1...
  • Page 94 If a LoadSure element has been selected then the tube size is displayed as pressure and bore. This screen also allows you to select the tube material used. Use the ∧ /∨ keys to move the selection bar over Tube material and press SELECT. Use the ∧...
  • Page 95 The PUMPHEAD MODEL screen allows the tube Lot Number to be recorded for future reference. Use the ∧ /∨ keys to move the selection bar over Tube lot number and press SELECT. Use the ∧ /∨ keys to scroll through the available characters for each digit. The available characters are 0-9, A-Z, and SPACE.
  • Page 96 Press CONFIRM followed by RE-CONFIRM to restore the defaults. Language Select language from the general settings menu to choose an alternative display language for the pump. The pump must be stopped before changing the language. Use the ∧ /∨ keys to move the selection bar to your required language. Press SELECT to confirm. m-730un-en-08.1...
  • Page 97: Control Settings

    Your selected language will now be displayed on screen. Press CONFIRM to continue, all displayed text will now appear in your chosen language. Press REJECT to return to the language choice screen. MODE menu Selecting MODE menu from the main menu will navigate you to access the sub-menu shown below. This is the same as pressing the MODE key.
  • Page 98 Applying a speed limit automatically re-scales the analog speed control response. Calibrated 4-20mA Calibrated 4-20mA user_max_flow user_max_flow recalibrated recalibrated Reset run hours Select Reset run hours from the control settings menu. Select RESET to zero the run hours counter. The run hours counter can be viewed by pressing INFO from your home screen.
  • Page 99: Configure Outputs

    21.4 Configure outputs Select Configure outputs from the CONTROL SETTINGS menu. Outputs 1 to 4 Use the ∧ /∨ keys and press SELECT to choose which output to configure. Use the ∧ /∨ keys and press SELECT to choose which pump status you require for the chosen output. The tick symbol indicates the current setting.
  • Page 100 Use the ∧ /∨ keys and press SELECT to choose the logic state of the chosen output. Press SELECT to program the output or BACK to cancel. 4-20mA output Select 4-20mA to configure the pump's 4-20mA output response. m-730un-en-08.1...
  • Page 101: Configure Inputs

    Use the ∧ /∨ keys and press SELECT to choose the required setting: Full scale 0 to 125 rpm — The 4-20mA output will be based on the full speed range of the pump. At 0 rpm the pump will output 4mA, at 125 rpm the pump will output 20mA. Match input scale —...
  • Page 102 Press SELECT to program the output or BACK to cancel. NOTE: On this model, Inputs 4 and 5 are configured for a pressure sensor. Disable remote stop in manual mode Users can disable/enable the remote stop input when the pump is in manual mode by using the following sequence to configure the 'start/stop' settings.
  • Page 103 The default is û. The start/stop input is not disabled in manual mode. Press SELECT to change the setting to ü. Press HOME to return and store the setting. The input is now disabled in manual mode. m-730un-en-08.1...
  • Page 104 Enable remote stop in manual mode The setting isü. The start/stop input is disabled. Press SELECT to open the logic state menu. Use the ∧ /∨ keys and press SELECT to choose the logic state of the chosen input for your connected control hardware.
  • Page 105: Help

    Press HOME to return and store the setting. The input is now enabled in manual mode. 21.6 Help Help Select Help from the main menu to access the help screens. m-730un-en-08.1...
  • Page 106: Troubleshooting

    Troubleshooting If the pump display remains blank when the pump is switched on, make the following checks: • Check that the mains power is available to the pump. • Check the fuse in the wall plug if one is present. •...
  • Page 107: Leak Detection

    22.1 Leak detection If a Watson-Marlow leak detector has been fitted to the pumphead, and a leak is detected; the pump will display the following message: Follow the instructions in "Tube replacement " on page 113 to replace your tube or tubing element.
  • Page 108: Technical Support

    Communication error (internal Attempt to reset by switching power OFF/ON. Err50 pump communications error and Or seek support. not a network error) 22.3 Technical support Watson-Marlow Fluid Technology Group Falmouth, Cornwall TR11 4RU Contact your local Watson-Marlow representative for support. www.wmftg.com/contact m-730un-en-08.1...
  • Page 109: Drive Maintenance

    Drive maintenance There are no user serviceable parts inside the pump. Please contact your local Watson-Marlow representative to arrange repair. m-730un-en-08.1...
  • Page 110: Drive Spares

    Drive spares Description Part No. Replaceable main fuse, type T5A, H 250V 20mm (Pack of 5) MRA3083A Foot (Pack of 5) MNA2101A Module Seal MN2516B Module switch cover MN2505M Gland (Std) GR0056 Glands (EMC) GR0075 Sealing washer for blanking plug or gland GR0058 Snap-Fit vent MN2513B...
  • Page 111: Pumphead Replacement

    Pumphead replacement Always isolate the pump from the mains power supply before opening any guard or track, or performing any positioning, removal or maintenance operation. 25.1 Pumphead replacement 520R 720R, 720RX, 720RE, 720REX safeguarding Primary safety on 730 series pumps is provided by the tool-lockable pumphead track.
  • Page 112 Refitting m-730un-en-08.1...
  • Page 113: Tube Replacement

    Tube replacement Always isolate the pump from the mains power supply before opening any guard or track, or performing any positioning, removal or maintenance operation. 26.1 Continuous tubing 720R 26.2 Tube elements 720RE m-730un-en-08.1...
  • Page 114 General guide to cleaning with solvents Chemical Cleaning precautions Aliphatic hydrocarbons Remove guard. Minimize rotor cap and clutch boot exposure to less than one minute (risk of attack). Aromatic hydrocarbons Remove guard. Minimize rotor cap and clutch boot exposure to less than one minute (risk of attack). Ketone solvents Remove guard.
  • Page 115: Ordering Information

    Ordering information 27.1 Pump part numbers • • • • • Model Plug options 3: S Ingress Protection Pumphead U: UK mains plug 4: U N: IP66 / NEMA 4X 00: 720R E: EU mains plug 5: Du S: SCADA IP66 / NEMA 4X* E0: 720RE A: American mains plug 6: Bp...
  • Page 116: Tubing And Element Part Numbers

    27.2 Tubing and element part numbers Continuous tubing for 720R pumpheads Pumpsil inch Marprene Bioprene silicone 902.0096.048 933.0096.048 913.A096.048 902.0127.048 933.0127.048 913.A127.048 12.7 902.0159.048 933.0159.048 913.A159.048 15.9 902.0190.048 933.0190.048 913.A190.048 19.0 902.0254.048 933.0254.048 913.A254.048 25.4 STA-PURE inch Neoprene Series PCS 961.0096.048 920.0127.048 961.0127.048...
  • Page 117 Sanitary elements with PVDF Tri-clamp style connectors STA-PURE Pumpsil inch Bioprene TL Series PCS silicone 961.0127.PFT 933.0127.PFT 12.7 913.A127.PFT 961.0159.PFT 933.0159.PFT 15.9 913.A159.PFT 961.0190.PFT 933.0190.PFT 19.0 913.A190.PFT 961.0254.PFT 933.0254.PFT 25.4 913.A254.PFT Industrial elements with PP Cam-and-Groove connectors Pumpsil inch Marprene TL Neoprene silicone 902.0127.PPC...
  • Page 118: Pumphead Spares

    27.3 Pumphead spares Continuous tubing models 720R and 720RX Number Spare Description MRA3062A Rotor assembly (720R) MRA0036A Rotor assembly (720RX) MRA0104A Knob assembly (4.8mm wall tubing) CN0090 Coupling half MR0880C Tube clamp MRA3061A Foot assembly CN0229 M12 blanking plug CN0088 Coupling spider MRA0027A Pivot pin assembly...
  • Page 119 Number Spare Description MR0662T Stud (Set to 61mm) MRA3063A Track assembly CN0228 M25 blanking plug MR0882M Eccentric bush MR3041T M8 x 307mm bolt (720RX) MR3040T M8 x 157mm bolt (720R) m-730un-en-08.1...
  • Page 120 LoadSure element models 720RE and 720REX Number Spare Description MRA3062A Rotor assembly (720RE) MRA0036A Rotor assembly (720REX) MRA0319A Knob assembly (4.8mm wall tubing) CN0090 Coupling half MR1118T Sliding clamp MRA3061A Foot assembly CN0229 M12 blanking plug CN0088 Coupling spider MRA0027A Pivot pin assembly MRA0034A Pivot pin assembly (720REX)
  • Page 121 Number Spare Description MR0662T Stud (Set to 61mm) MRA3064A Track assembly CN0228 M25 blanking plug MR0882M Eccentric bush MR3041T M8 x 307mm bolt (720REX) MR3040T M8 x 157mm bolt (720RE) m-730un-en-08.1...
  • Page 122: Performance Data

    Performance data 28.1 720R, 720RE, 720R/RX and 720RE/REX performance data Pumping conditions All performance figures in this operating instruction have been recorded against peak pipeline pressures. This pump is rated to 2 bar (30psi) peak pressure when fitted with a 720R, 720RE, 720R/RX or 720RE/REX pumphead using high-pressure tubing.
  • Page 123 pressure—psi pressure—psi pressure—psi pressure—psi 2000 1800 9.56mm 1600 1400 12.7mm 1200 15.9mm 1000 19.0mm 25.4mm 0.25 0.50 1.00 1.50 2.00 0.53 0.40 0.27 0.13 pressure— mmhg pressure—mmhg pressure—bar pressure—bar Pressure Suction (-) bar (+) bar m-730un-en-08.1...
  • Page 124 720R/RX and 720RE/REX flow rates 730 cased pumps’ performance limits 0.25 bar (3.6psi) 0.5 bar (8psi) 1 bar (15psi) 1.5 bar (22psi) 2 bar (30psi) Twin pumphead (720R/RX, Max flow 720RE/REX) speed L/hr speed speed speed speed flow flow flow flow (rpm)* (USGPH)
  • Page 125: Trademarks

    Trademarks Watson-Marlow, LoadSure, Qdos, ReNu, LaserTraceability, Pumpsil, PureWeld XL, Bioprene, Marprene are registered trademarks of Watson-Marlow Limited. Tri-Clamp is a registered trademark of Alfa Laval Corporate AB. STA-PURE Series PCS and STA-PURE Series PFL are trademarks of W.L.Gore and Associates.
  • Page 126: Disclaimers

    Disclaimers The information contained in this document is believed to be correct but Watson- Marlow Fluid Technology Group accepts no liability for any errors it contains and reserves the right to alter specifications without notice. WARNING: This product is not designed for use in and should not be used for, patient- connected applications.
  • Page 127: Publication History

    Publication history All versions updated, combined to single source and unified to Issue 4 09.18 m-730un-gb-01 730 UN pump First published 01.17 m-730en-01 530 EN pump First published 04.2020 m-730un-en-08.1...

This manual is also suitable for:

730 un720r720re720rx720rex

Table of Contents