Cleaning The Sample Filtration And Sample Feed; Mechanical Cleaning Of The Filter Plate - Xylem YSI Alyza IQ PO4 Operation Manual

1- and 2-channel measuring systems for online determination of ortho-phosphate in aqueous samples
Hide thumbs Also See for YSI Alyza IQ PO4:
Table of Contents

Advertisement

Alyza IQ PO4
Terminating the
maintenance
routine
ba76201d03
45
After maintenance activities at the measuring unit:
Close the front cover of the measuring unit and the doors of the
Alyza IQ.
46
Carry out the Messung vorbereiten function.
The temperature control of the Alyza IQ is started.
Wait for the temperature of the Alyza IQ to be regulated (display in the
tab Status).
47
Fill the tubes (Alyza menu / tab Wartung / Manuelle Funktionen / Sys-
tem befüllen or Befüllen (einzeln))
48
Calibrate the measuring system with the new replacement parts
(Alyza menu / tab Wartung / Manuelle Funktionen / Kalibrieren (2-
Punkt))
49
Carry out the Alyza IQ STARTEN function. Measurement is started and
the measured value is displayed in the measured value display after
approx. 5 ... 7 minutes.
50
Carry out the Wartungszustand ausschalten function.
5.6

Cleaning the sample filtration and sample feed

Cleaning of the filter module, see the following documents:
 Mounting instructions for the filter module (FM/PC)
 Alyza IQ - short instructions for cleaning the filter membrane
module (included in the scope of delivery of the Alyza IQ )

5.6.1 Mechanical cleaning of the filter plate

The time when to clean the filter plate can be determined empirically based on
the negative pressure display on the filtration pump.
 A pulsing negative pressure manometer suggests a normal operating condi-
tion of the filtration unit.
 With a delivery height of approx. 2 m and a new filter plate, the negative pres-
sure is approx. -0.3 bar.
Every meter of delivery height increases the negative pressure by approx. -
0.1 bar.
If the negative pressure increases by a further -0.3 ... -0.4 bar with time, the
filter plate is covered with solid matter from the sample and has to be cleaned.
 If, after the mechanical cleaning, there is still negative pressure, a chemical
cleaning is required, or the intake line may be blocked.