Section 4 Operation

Editor - Setpoints

Editor Setpoints
is used to view/adjust the thermostat's five Setpoints and Real Temperature
Adjustments (RTA). The display shows the temperature measured by the internal or external sensor, if
this temperature does not accurately reflect the actual temperature in the reservoir an RTA is required.
The RTA can be set ±10°C (±18°F).
As an example, if the thermostat temperature is stabilized and displaying 20°C but a calibrated reference
thermometer reads 20.5°C, the RTA should be set to -0.5°C. After you enter a RTA value allow unit to
stabilize before verifying the temperature in the bath. NOTE If display accuracy is required, we recom-
mend repeating this procedure at various setpoint temperatures and on a regular basis.
NOTE You cannot adjust the setpoint closer than 5°C to either of the fluid's system limits, see Fluids
Type in this Section, or beyond the bath's temperature range, see Section 2. For example, an A10 with
glycol-water cannot be set lower than -10°C or higher than +95°C.
1. Use the arrow buttons to highlight
SP1
Editor
Settings
System
Installation
Setpoints
3. With
highlighted, press
display the list. NOTE Use the down arrow to
SP5
display
.
SP1 RTA
SP2 RTA
SP3 RTA
SP4 RTA
Menu
5. With the desired line highlighted press
A highlight will appear on the last digit.
Setpoint
RTA int
RTA ext
Menu
4-8
Editor.
Menu
to
.
x
+xxx.x
+xx.xx
+xx.xx
2. Press
to display the submenu.
Setpoints
Ramp Program
Auto start
Menu
4. The Setpoint and RTA are changed using the
same procedure. With the desired setpoint high-
lighted press
to display the submenu.
Setpoint
RTA int
RTA ext
Menu
6. Use the left and right arrows to move the high-
light to the desired digit and then use the up and
down arrows to change the value. Once all the
desired changes are made, press
the changes.
+xxx.xx
+xx.xx
+xx.xx
to save