Section 3 Installation

Filling Requirements

CAUTION
CAUTION
CAUTION
3-12
Ensure the reservoir drain port on the front of the unit is closed and
that all plumbing connections are secure. Also ensure any residue is
thoroughly removed before filling the unit.
Before using any fluid refer to the manufacturer's MSDS and EC
safety data sheets for handling precautions.
To avoid spilling, place your containers into the bath before filling.
With a low level WARNING the unit continues to run, with a FAULT
the unit will shut the refrigeration, pump and heater will shut down, see
Section 7. The low level warning is at approximately 3 cm (1 1/4") below
the top, the low level fault is at approximately 7.5 cm (3").
Avoid overfilling, oil-based fluids expand when heated.
When pumping to an external system, keep extra fluid on hand to
maintain the proper level in the circulating lines and the external system.
NOTE Monitor the fluid level whenever heating the fluid.

Draining

Before draining any fluid refer to the manufacturer's MSDS and EC
safety data sheets for handling precautions.
Ensure the fluid is at a safe handling temperature, ~55°C. Wear
protective clothing and gloves.
place a suitable vessel underneath the drain. If desired, attach an
8 mm id tube on the drain.
slowly
Turning the drain cap more than 1 1/2 turns will result in the drain
cap and fitting coming off the bath.
In this case, the drain fitting can be screwed back into the unit. Attaching
the cap onto the fitting will aid in installation. If required, contact us for
additional information.
turn the drain plug until flow is observed.
Installed Drain Fitting with Cap Removed