Changing The Setpoint - Thermo Scientific ADVANCED Series Manual

Thermostats, refrigerated bath circulators, heating bath circulators
Hide thumbs Also See for ADVANCED Series:
Table of Contents

Advertisement

Section 4 Operation
Changing the
Setpoint
Setpoint Value Window
Setpoint Value Window
4-4
NOTE You cannot adjust the setpoint closer than 5°C to either of
the fluid's system limits, see Fluids Type in this Section, or beyond
the bath's temperature range, see Section 2. For example, an A10 with
glycol-water cannot be set lower than -10°C or higher than +95°C.
The Setpoint is the desired fluid temperature. The thermostat can store
SP1
up to five setpoints,
the stored setpoint values is discussed later in this Section.
Use the navigation arrows and move to the
press
to highlight it as shown below.
SP1
+020.00
Internal Temp..
Use the up and down navigation arrows to bring up the desired setpoint
and then press
.
The display on the Setpoint Value Window will now indicate the corre-
sponding setpoint's stored value.
SP2
+035.00
Internal Temp..
If desired, you can change the displayed setpoint value by using the
navigation arrows to highlight the Setpoint Value Window and then
pressing
. The right-most digit will highlight and flash.
+020.00
Use the left and right arrows to move the highlight to the desired digit
and then use the up and down arrows to change the value. Once all the
desired changes are made, press
NOTE Using this procedure also changes the setpoint's stored value.
NOTE The setpoint can be changed with the unit running.
SP5
through
. The procedure for changing
SP1
MENU
+024.29 °C
MENU
+024.29 °C
to save the change.
window and then
In
in
Thermo Scientific