96 PRODIGIT
5.10.1.7. Setting OPP VTH 0.600V,press OPP key to the next step.
5.10.1.8. Press START/STOP Test key.
5.10.1.9. the UUT's output voltage drop-out lower than the threshold voltage (V-th
5.10.2 Remote control OPP
EX:
REMOTE
TCONFIG OPP
OPP:START 3
OPP:STEP 1
OPP:STOP 5
VTH 3.0
WL 0
WH 5
NGENABLE ON
START
TESTING?
NG?
OPP?
STOP
5-11. SHORT testing
setting), and the OPP trip point is between W_Hi and W_Lo limitation, then
lower 5 digits LCD display will shows "PASS", otherwise shows "FAIL".
(Set Remote)
(Set OCP test)
(Set start load watt 3W)
(Set step load watt 1W)
(Set stop load watt 5W)
(Set OPP VTH 3.0V)
(Set watt low limit 0W)
(Set watt high limit 5W)
(Set NG Enable ON)
(Start OPP testing)
(Ask Testing? 1:Testing,0:Testing End)
(Ask PASS/FAIL?,0:PASS,1:FAIL)
(Ask OPP watt value)
(Stop OPP testing)