Magnamed OxyMag Operation Manual page 75

Transport and emergency ventilator
Hide thumbs Also See for OxyMag:
Table of Contents

Advertisement

Problem
3. O2 Pressure Sensor Failure
1. Inspiratory or expiratory branch obstructed
Obstruction alarm
2. Patient's airway obstruction
1. Leak in patient circuit
Low Maximum Pressure
alarm
2. Inspiratory pressure lower than set alarm limit
1. Expired volume is greater than the set alarm limit.
High Volume alarm
2. Flow sensor is out of calibration.
1. Expired volume is less than the set alarm limit.
Low Volume alarm
2. Flow sensor is out of calibration.
1. FiO2 adjust is above alarm limit;
High FiO
alarm
2. O2 cell is out of calibration;
2
3. Damaged O2 cell.
1. FiO2 adjust is below alarm limit;
Low FiO
alarm
2. O2 cell is out of calibration;
2
3. Damaged O2 cell.
1. Expired CO2 is above alarm limit
High EtCO
alarm
2. Capnography sensor is out of calibration
2
3. Capnography sensor damaged.
1. Expired CO2 is below alarm limit
Low EtCO
alarm
2. Capnography sensor is out of calibration
2
3. Capnography sensor damaged.
CO
i alarm
1. Inspired CO2 is above alarm limit
2
75
Possible Causes
Solutions
2. Connect the O2 hose to the device.
3. Contact Magnamed Technical Assistance.
1. Clear it;
2. Clear or aspirate the patient's airway.
1. Locate the leak and correct;
2. Increase the lower limit of the pressure alarm
or increase the set inspiratory pressure.
1. Change the parameters set in ventilation
mode or set the upper limit of the volume
alarm.
2. Perform self-test.
1. Change the parameters set in ventilation
mode or adjust the lower limit of the volume
alarm.
2. Perform self-test.
1. Change the set FiO2 or upper alarm limit;
2. Calibrate O2 cell;
3. Contact Magnamed Technical Assistance.
1. Change the set FiO2 or lower alarm limit;
2. Calibrate O2 cell;
3. Contact Magnamed Technical Assistance.
1. Change the parameters set in ventilation
mode or set the upper limit of the EtCO2 alarm.
2. Calibrate capnography sensor.
3. Contact Magnamed Technical Assistance.
1. Change the parameters set in ventilation
mode or set the internal limit of the EtCO2
alarm.
2. Calibrate capnography sensor.
3. Contact Magnamed Technical Assistance.
1. Change the parameters set in ventilation
mode or set the upper limit of the CO2i alarm.
OxyMag_rev26