Cpap/Psv - C; Ressure Controlled Cycle 60 Ontinuous Pressure Upport - Magnamed OxyMag Operation Manual

Transport and emergency ventilator
Hide thumbs Also See for OxyMag:
Table of Contents

Advertisement

8.6 CPAP/PSV – Continuous Pressure Ventilation with Pressure
Support
Description::
In this mode, the patient breathes spontaneously on a continuous
positive pressure and breathing is assisted by a Pressure Support
(ΔPS). Usually when analyzing the flow curve, a flow peak is seen
which decreases over time.
Cycling occurs by flow, is adjustable between 5% and 80% peak
inspiratory flow measured.
If the Pressure Support (ΔPS) value is set to 0 (ZERO) or both cycle
trigger ways (pressure or flow) are disabled, pure CPAP mode will be
activated, i.e., without pressure support. In this condition, PEEP
parameter will be displayed as CPAP.
62
Figure 25: PSV (CPAP + ΔASB) Curves
Set Parameters:
PEEP or CPAP;
ΔPS (Pressure Support – PEEP);
TRIGGER BY FLOW;
TRIGGER BY PRESSURE;
FiO2
CYCLING BY FLOW (% FLOW);
RISE TIME
BACK-UP MODE (VCV,PCV, PLV-
NEONATAL or WITHOUT BACK-UP)
RATE (back-up VCV, PCV and PLV);
RATIO I:E (back-up VCV and PCV);
MAXIMUM PRESSURE (back-up VCV);
VOLUME (back-up VCV);
PAUSE (back-up VCV);
FLOW WAVEFORM (back-up VCV);
INSP PRESSURE (back-up PCV and PLV);
INSPIRATORY TIME (back-up PLV);
FLOW ( - back-up PLV)