Magnamed OxyMag Operation Manual page 70

Transport and emergency ventilator
Hide thumbs Also See for OxyMag:
Table of Contents

Advertisement

MEDIUM PRIORITY
High Minute volume
Low Minute volume
High Rate
Low Rate
High PEEP
Low PEEP
CO
out of range
2
IRMA out of range
IRMA zero required
Hight Temperature
Low Temperature
High FiO
2
Low FiO
2
70
Delay Time
It is triggered when the minute volume of the patient has exceeded the alarm
< 3 cycles
value set as upper limit of minute volume
It is triggered when the minute volume of the patient is below the alarm value
< 3 cycles
set as lower limit of minute volume
It is triggered when the respiratory rate of the patient has exceeded the alarm
< 3 cycles
value set as upper limit of the respiratory rate
It is triggered when the respiratory rate of the patient is below the alarm value
< 3 cycles
set as lower limit of the respiratory rate
It is triggered when the pressure in the end of exhalation (PEEP) has exceeded
< 3 cycles
the alarm value set as upper limit of PEEP
It is triggered when the pressure in the end of exhalation (PEEP) is below the
< 3 cycles
alarm value set as lower limit of PEEP
It is triggered when IRMA CO
< 3 seconds
It is triggered in one of the situations with the CO
< 3 seconds
It is triggered when there is need of Zero calibration of IRMA CO
< 3 seconds
It is triggered when the environment condition is above 50ºC.
< 3 seconds
It is triggered when the environment condition is below -18°C.
< 3 seconds
It is triggered when the measured FiO
< 3 cycles
upper limit of FiO
It is triggered when the measured FiO
< 3 cycles
lower limit of FiO
Description
reading is incorrect
2
Internal temperature of operation is out of scale or
Ambient pressure of operation is out of scale
2
2
2
2
sensor:
2
sensor
2
has exceeded the alarm value set as the
has exceeded the alarm value set as the
OxyMag_rev26