Magnamed OxyMag Operation Manual page 41

Transport and emergency ventilator
Hide thumbs Also See for OxyMag:
Table of Contents

Advertisement

Attribute
Breathing Detection
Respiratory Rate
Fi and ET
Waveform
Diagnostic Parameters
Information
Method to calculate gas level reading
Sensor
Data Acquisition Rate
Data Sampling Rate
Calibration
41
Adds less than 6ml dead space;
Pressure drop lower than 0.3cmH
Pediatric (Disposable):
Adds less than 1ml dead space;
Pressure drop less than 1.3cmH
OUTPUTS
Adaptive threshold, minimum 1% volume change in CO
0 - 150bpm. Respiratory Rate is shown every 3 breaths and the
average value is updated every breath.
Fi and ET are shown after a breath and their averages are continually
updated.
IRMA CO
: CO
.
2
2
IRMA CO
: CO
.
2
2
Atmospheric Pressure, review of software and hardware, serial
number.
Detection of New Breath, Apnea, Check Adapter, Accuracy Not
Specified and Sensor Error.
The highest CO
concentration during a respiratory cycle with a weight
2
function applied to favor values closer to the end of the cycle
CO
Gas Analyzer:
2
Gas analyzer with 2 - 9 NDIR (Non-Dispersive Infrared) Channels
which measures in the range of 4 - 10µm. Makes correction of
pressure, temperature and interference across the spectral range.
10 kHz
20 Hz
Resetting recommended at each replacement of Airway Adapter. No
need for specific infrared calibration.
Specification
0 at 30LPM.
2
0 at 10LPM.
2
concentration.
2
OxyMag_rev26