To return to the screen with the button for selection of
charts, data, settings and alarms press button
automatically adjust the alarm values select the parameter
"Automatic" and choose the default limit: OFF, 10%, 20%
or 30%. The alarm limits of ventilator parameters
(Pressure, PEEP, MV, Volume, FiO2 and Frequency) will
be automatically adjusted:
a) In the lower limit: for the value of the parameter currently
measured minus the percentage selected in automatic;
b) In the upper limit: for the value of the parameter
currently measured plus the percentage selected in
automatic;
c) If OFF selected, these alarms return to the alarm default
values for the type of patient set at startup;
To enable the automatic adjustment, it is required that the
ventilator is not in STAND-BY.
12. By pressing SETTING button, you can:
Change the height of the Patient thereby defining the ideal
weight (IMC 22), recalculating the default ventilatory
parameters of this patient;
Turn on or off the NIV (Non Invasive Ventilation) using a
mask. When NIV is activated, there will be compensation
of leaks of approximately 40L.min-1, depending on the
ventilator settings;
Turn on or off the extern blender compensation. In this
condition the device don't allow the FiO
adjustment will be done directly in the blender
1 With extern Blender it is possible to adjust the FiO
Note
It is not necessary to discard calibration gases.
32
Turn on or off the sigh function. In this condition, there will
be a sigh every 100 cycles in control modes.
. To
Select the language of the equipment;
Press O2/CO2 tab to perform calibration of O2 and CO2
meters. Press button "Calibrate FiO2" to calibrate the
oxygen sensor. Press button "Calibrate CO2" to calibrate
CO2 sensor.
Press Ventilator tab to:
Set:
View:
13. Press the screen on chart area and menu. A panel of
buttons will be displayed for selection of charts, data,
settings and alarms. If no external sensor is connected to
the equipment (capnography or oximeter) the respective
button will not appear.
. The FiO
2
2
;
1
from 21 to 100%.
2
Audio volume – to set the alarm audio volume.
o
Use the button "knob and confirm" to make this
adjustment. This parameter always starts in the
maximum level 5.
Pressure unit – Press the desired unit;
o
Data of the last test performed: breathing
o
system leakage, compliance and resistance
done in initial tests;
Total hours of use of the equipment;
o
Hours past since the last maintenance;
o