Magnamed OxyMag Operation Manual page 117

Transport and emergency ventilator
Hide thumbs Also See for OxyMag:
Table of Contents

Advertisement

The Oxymag cabinet interface has been tested as specified in the table below using the test methods
Band
[ MHz ]
380 to 390
430 to 470
704 to 787
800 to 960
1.700 to 1.990
2.400 to 2.570
5.100 to 5.800
D) Electrical Safety
The following are the precautions that should be observed when combining these items (non-medical equipment) with the system.
• Items
that
requirements of standard IEC 60601-1
cannot be placed within 1.5 from the
patient.
• All items (electromedical or non-
medical equipment) connected to the
system with input/output signal cable
must receive power from an AC source
using separate transformer (according
to standard IEC 60989) or provide
additional earth protective conductor.
• Multiple portable sockets with switch
used on AC power supplies must
comply with IEC 60601-1-1 and cannot
be installed on the floor. Do not use
more than one power strip with
portable switch.
• Do not connect directly the non-
medical electrical equipment to an AC
wall outlet. Use AC power supply with
117
Fields in the vicinity of RF wireless communication equipment
specified in IEC 61000-4-3
Freq. test
[ MHz ]
385
450
710
745
780
810
870
930
1.720
1.845
1.970
2.450
5.240
5.500
5.785
do
not
meet
Modulation
Pulse, 18 Hz
FM, 1 kHz,
Deviation from ± 5kHz
Pulse, 217 Hz
Pulse, 18 Hz
Pulse, 217 Hz
Pulse, 217 Hz
Pulse, 217 Hz
its own transformer. Otherwise, the
the
leakage current will increase above the
levels accepted by IEC 60601- under
normal conditions and single-failure
conditions. This may cause dangerous
electrical shock to the patient or
operator.
• After connecting any equipment into
these outlets, the system should
undergo a complete test for leakage
current (according to standard IEC
60601-1).
• The operator of the electromedical
system must not touch the non-
medical electrical equipment and the
patient at the same time. This may
cause dangerous electrical shock to
the patient or operator.
Trial level
[ V/m ]
27
28
9
28
28
28
9
OxyMag_rev26