Table 10: List of parameters in the Neonatal mode
PLV
Pinsp
Rate
Tinsp
PEEP
Flow
FiO
2
Flow Trigger
Pressure Trigger
5.
By pressing key PEDIATRIC the ventilator will start
ventilation with the following parameters:
Table 11: List of parameters in the Pediatric mode
PCV
Pinsp
Rate
Ratio I:E
PEEP
FiO
2
Flow Trigger
Pressure Trigger
Rise Time
6.
By pressing the key ADULT, the ventilator will start
ventilation with the following parameters:
Table 12: List of the parameters in the Adult mode
VCV
Vt
Rate
Ratio I:E
31
Default
15 cmH2O
40 min
-1
0,5s
5 cmH2O
6 L.min
-1
40%
OFF
OFF
Default
15 cmH2O
22 min
-1
1:2
5 cmH2O
50%
OFF
OFF
0,1s
Default
350 mL
17 min
-1
1:2
VCV
PEEP
Pmax
Pause
FiO
2
Flow Trigger
Pressure Trigger
Wave Flow
7.
After the startup sequence, the equipment will display the
chart screen of the ventilator. Audible alarm will be
disabled in the first 2 minutes. Note that the white bar next
to the alarm silencing symbol is reduced over time. After 2
minutes the audible alarm will be activated.
8.
Press the ventilation mode button
the following screen to select the mode.
9.
Press the desired mode button and after confirmation of
adjustment parameter required for this ventilation mode
and the ventilation will immediately start.
10. To change a parameter, press the corresponding button.
The parameter will become YELLOW indicating that is
selected, allowing the change. Turn the knob clockwise to
increase the value and counterclockwise to reduce. To
confirm the change, press the knob and confirm or tap on
the parameter button to be set on screen to activate the
new value.
11. Press ALARM button
screen will appear, the figures below follow the order of
the conditions described in the previous item. Tap on
alarm to be set and use the knob and confirm to set the
value. When the desired value is set confirm pressing the
knob and confirm.
Default
5 cmH2O
35 cmH2O
30%
50%
OFF
OFF
Square