Magnamed OxyMag Operation Manual page 104

Transport and emergency ventilator
Hide thumbs Also See for OxyMag:
Table of Contents

Advertisement

Item
Maximum
1
2
Total Volume
3
Minute volume
4
Time for Apnea
5
Respiratory Rate
6
7
Automatic limit
8
9
Heart Frequency ²
10
11
12
CO2 inspired ²
¹ Every time the equipment starts up or there is a change of patient type or the battery power runs out without plugging the
ventilator to mains, the alarms will assume the default values indicated for each type of patient.
Alarms available only with the use of optional external sensors.
2
Only to be applied to the alarms related to basic ventilation parameters (Maximum Pressure, PEEP, Minute volume and
3
Respiratory Rate).
104
Alarm
Setting
OFF; 0 to 80
Pressure
PEEP
OFF; 0 to 40
OFF; 0 to 3000
OFF; 0 to 99
OFF; 5 to 60
Alarm
OFF; 0 to 150
FiO2
OFF; 35 to 100
OFF, 10, 20 and
setting ³
30
OFF;25 to 240
SpO
²
OFF; 1 to 100
2
EtCO
²
OFF; 0 to 80
2
OFF; 0 to 80
Standard Alarm¹
Limit
NEO
PED
High
30
30
Low
OFF
OFF
High
10
15
Low
OFF
OFF
High
50 mL
500 mL
Low
OFF
OFF
High
5.0
10
Low
OFF
OFF
High
OFF
OFF
Low
10
10
High
60
60
High
OFF
OFF
Low
80
80
High
OFF
Low
180
120
High
80
40
Low
85
85
High
OFF
OFF
Low
45
45
High
OFF
OFF
---
4
4
Unit
ADU
40
cmH
O
2
OFF
20
cmH
O
2
OFF
1.0 L
mL
OFF
20
OFF
L
OFF
15
s
60
min
-1
OFF
80
%
%
120
bpm
40
85
%
OFF
45
mmHg
OFF
4
mmHg
OxyMag_rev26