Dräger NAD Operator's Instruction Manual page 16

Omicron monitor
Table of Contents

Advertisement

3
Operating Instructions
Table 3-2. Middle Alert Priority Alarms
Display
APNEA VOL
MINUT HI
TIDAL LOW
MINUT LOW
Table 3-3. Low Alert Priority Alarms
Display
O2 HI LMT
O2 SENSOR
Saving Alarm Limits
When the monitor leaves the manufacturer's production facility, each
alarm limit is set to the values listed in
Each time the NAD Omicron Monitor is energized, the alarm limits are set
to the last saved default values. Use this procedure to set the default
values to the current alarm limit settings:
1. Select the parameter to be changed and depress the parameter keys.
2. Change the setting by pressing the high or low limit button and the up
3. Depress the parameter keys again (this will cause the numbers on the
4. Simultaneously press the up and down arrow keys until the monitor
Even when powered off, the monitor retains stored alarm limits.
NOTE: If the alarm limit settings for Tidal Volume and Minute Volume
3-6
Designation
Apnea Volume > 15 sec
Minute Volume High Limit
Tidal Volume Low Limit
Minute Volume Low Limit
Designation
O
High Limit
2
O
Sensor Failure
2
or down arrow keys.
monitor to disappear).
displays the message SAVE DATA; the alarm limit settings are saved.
are not saved as described above, depressing the Neonate/Adult
key will restore those alarm limits to the last user-saved or
factory-saved preset alarm parameters (see
for more information).
RETURN TO CD-ROM TABLE OF CONTENTS
Description
Tidal Volume is less than 30 mL for
more than 15 seconds.
Minute Volume Value is above the
High Limit Set Value.
Tidal Volume Value is below the Low
Limit Set Value.
Minute Volume Value is below the
Low Limit Set Value.
Description
O
concentration is above the High
2
Limit Set Value.
O
Sensor failed or disconnected.
2
Table 3-4
(factory presets).
Table 3-4 on page 3-7