Performance Test for Transducers

Performance Test for Transducers

Error Handling

22
This tests the signal path to and from the transducers.
1. Turn on the monitor and the recorder or printer.
2. Connect one or two US cables and the TOCO transducer.
3. Check that the display acknowledges the connected transducers in the
appropriate frames.
Also refer to "Testing Transducers" on page 90.
There are two types of error that you may see: those that the monitor cannot
self-detect, and those that it can self-detect.
• If an operational error occurs that is not self-detected, the monitor shuts
down. There will be no audible or visual warnings or indicators.
• If a self-detectable error occurs, the monitor will do the following:
– Display the error screen
– Sound an error tone
– Discontinue all monitoring functions
Chapter 3 - Tests and Error Codes