Table 22 Enclosure Leakage Current - Philips FM-2 Service Manual

Antepartum fetal monitor
Table of Contents

Advertisement

Test Steps
Chapter 8 - Test and Inspection Matrix
1. Connect the external power supply output cord to the monitor. Then
connect the appropriate external power supply input power cord to the
analyzer as recommended by the analyzer operating instructions.
2. Using the appropriate test cable, connect the analyzer to a fixing nut of
the RS-232 connector on the back of the monitor.
3. Turn the monitor on.
4. Perform the test as recommended by the analyzer operating instructions.
The analyzer leakage current indication must not exceed the values listed in
Table 22.
Table 22
Enclosure Leakage Current
Test Condition
Normal
1
S.F.C.
(Open Supply)
1. S.F.C. = Single Fault Condition
Maximum Leakage Current
Polarity
IEC 601-1 ES1
Normal
Reversed
Normal
Reversed
Safety Testing
(µA)
AAMI/ANSI
Standard
100 µA
100 µA
100 µA
100 µA
500 µA
300 µA
500 µA
300 µA